Beckman Coulter - TOC PAT700
TOC Analyzer Liquid Particle Counters
Active Questions & AnswersAsk a Question
There are no current Discussions
TOC Analyzers Service ProvidersView All (11)
Documents & Manuals
There are no Documents or Manuals available.
Features of TOC PAT700
Have confidence in the data - assured with UV Detect, excursion alarming, and fully traceable calibration Implement the FDA PAT initiative for on-line release for TOC and Conductivity eliminating time consuming and costly laboratory analysis.
Reduce the risk of product recall and speed up water loop troublshooting when an excursion alarm occurs The PAT700 internally captures two samples and then re-analyzes the first sample to validate the excursion alarm and saves the second sample for further root-cause analysis.
Confirm production vessels are clean and ready to go on-line faster with Clean-in-Place (CIP) analysis The PAT700 clean-in-place analysis maximizes production up-time by providing immediate on-line results for both TOC and conductivity. Providing instant on-line verification that your production vessel is ready for production changeover.
Save hours and expand laboratory throughput with the PAT700 36 bottle auto-sampler Using the ASX-7200 auto-sampler allows laboratory users to set up 36 bottles for TOC analysis and walk away. Using the PAT700 there is no requirement for costly reagents or carrier gases.
General Specifications
There are no General Specifications available.
Additional Specifications
TOC
Operating Range
0.5 to 2,000 ppb as Carbon
Display Resolution
0.1 ppd
Accuracy
+/-1 ppd or +/-5% whichever is greater
Repeatability
+/-0.3 ppd or +/-1% whichever is greater
Limit of Detection
0.5 ppb
Maximum Input Conductivity
0.2 uS/cm for all waters, 1.0 uS/cm for all neutral waters, 5.0 uS/cm for water with CO2 as the sole conductive species
Conductivity
Conductivity
Range 0.05 to 150 uS/cm (@25* C)
Display Resolution
0.01 uS/cm
Conductivity Accuracy
+/-2% over full range (uncompensated)
Resistivity
Resistivity Range 0.2 to 18 MQ-cm (@25*C)
Display Resolution
0.01 over full range
Available Modes
Temperature compensated to 25*C, or uncompensated
Temperature
Ambient Operating Range
10 to 40*C (50 to 104*F)
Measurement Accuracy
+/-0.4*C
Sample Water Range
5 to 95*C (41 to 203*F)
Display Resolution
0.1 over full range
Physical Specs
UV Lamps
2, with UV Detect technology
Interface/Display
Color with touch screen
Maximum Altitude
4,000 m (13,125)
User I/O Wiring
Three, 3/4-inch conduit openings or quick disconnect fittings
Standards System
Onboard, Automated Standards Introduction System (OASIS)
Dimensions
59.7 w X 22.9 d X 25.4 h (23.5 X 9 X 10 inches)
Weight
13.6 kg (30 lbs)
Sample Inlet Flow Rate Range
60 mL/min to 300 mL/min
Sample Inlet Pressure Range
10 to 100 psi (69 to 690 kPa)
Compliance
Installation Category
II
Pollution Degree
2, IEC 61010-1
CE Compliance
EN 61010-1 and EN 61326
Safety Rating
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Enclosure Rating
Conduit version: IP56 Quick connect version: IP46
Release Tests
USP , USP , JP 16, EP 2.2.44
New Features
CIP
Selectable mode for Clean-In-Place analysis
Dual Stream option
Toggle or programmable stream switching
Excursion sampling
Minimum flow rate to fill excursion bottle = 160 mL/min
Rouge detection
Identifies oxidation cell contamination from roughing