Pharmatest USA - PTZ AUTO4EZ
Fully Automated Disintegration Testing
Depending which size of sample needs to be tested select either the PT-MKT or PT-MKT33 electronical baskets which are both equipped with a temperature sensor to monitor the disintegration buffer temperature, regardless of how many vessels there are in operation. The operational concept of the electronic basket types PT-MKT and PT-MKT33 are simple in design and operate through electrical contacts in each of the 6 (3) working positions of each basket. This is an easy to maintain device and offers the user the additional security that any of the electronic discs may be used in any of the disintegration tubes. There are no overly complicated detectors which need to be “tuned” to each particular tube or disk. Rigid tiny metal rings are used to make the contact of the disks with the sieve segments. Connect a common PCL printer to the parallel printer port of the PTZ AUTO EZ and at the end of a run, when all tablets are disintegrated, a full test report is printed including individual disintegration, maximum and minimum time. The standard instrument allows to edit maximum 16 testing methods which can be selected by authorised users to start a test.
Active Questions & AnswersAsk a Question
There are no current Discussions
Disintegration Tester Service ProvidersView All (9)
Documents & ManualsView All Documents
Features of PTZ AUTO4EZ
- alpha-numerical keyboard * 320x240 Pixel large LC Display * Test Information Filing up to 16 Methods * Individual start up for each station * Media Temperature measuring in each vessel * USP and EP compliant PT-MKT baskets * USP and EP PT-MKT33 baskets optional * Print out of Run-Time Log at end of test * Print out of Test Report incl. Statistics and Graphics * OQ/PQ information system * Manual Stop of Disintegration Times, press relevant key only, results are marked “M” * Automated detection of the Disintegration Time of each Sample, results are marked “A” * RS-232 interface, parallel Printer Interface - USB adapter option * ABS Auto-Basket-Stop at highest position by built-in Basket Lift * TSC system (Thermostat Safety Control system) * Control by 21CFR part 11 compliant Software PTZ32, alternatively standard software package
General Specifications
There are no General Specifications available.
Additional Specifications
Test Station:
4 independent operateable stations, holding each 1 PT-MKT or PT- MKT33 basket
Display:
LCD Display 320x240 Pixel
Keyboard:
alpha-numerical keyboard and function keys
OQ information: programmable time sequence information to perform OQ test
Security:
password protected instrument calibration and validation procedures
Stroke Frequency:
30 strokes / minute
Accuracy:
± 1 stroke
Stroke Height:
55 mm
Accuracy:
± 1 mm
Basket Lift:
rises the baskets out of the media at end of test run
Thermostat:
built-in 1000 W heating tube and circulation pump Temperature adjustable within 25°- 45.0° C
Accuracy:
± 0.3° C
Safety System:
pressure switch to protect the heater from overheating (no flow - no heating - no pumping)
thermo switch and thermo fuse to protect the heating system in case of an electronical fault
Memory:
up to 16 different test descriptive methods
Printout:
online printing of test report including individual disintegration times, statistics and graphics
Interface:
RS-232 port
Parallel PCL Printer port
USB adapter for both available as option