Anatel TOC analyzers are the industry standard for purified water and ultrapure water.
A separate, portable proving system, the CSV module, provides an easy to use system for performing validation, calibration, and system suitability testing. The Anatel TOC600 can be calibrated using multiple points, offering an accurate calibration curve. The self-contained CSV module can be used across multiple analyzers and requires less standards for a significant reduction in the overall cost of ownership.
Each Anatel TOC600 can be operated using remote control inputs. Up to eight Anatel TOC600 analyzers can be networked to one C80 controller providing a low cost multiple analyzer system. Operation of the CSV Module is directed through the C80 controller. The TOC600 automatically calculates results that are formatted into reports for display, print, and archive.
With a robust design, high-reliability, unmatched performance, USP and EP compliance, and a low cost of ownership, the Anatel TOC600 represents a new class in value for TOC analyzers.
Active Questions & AnswersAsk a Question
There are no current Discussions
TOC Analyzers Service ProvidersView All (11)
Documents & ManualsView All Documents
Features of ANATEL TOC600
- Plug and play - easy installation, start up, and operation
- Online TOC measurement with complete, accurate sample oxidation
- Meets USP , ,and EP 2.2.44 requirements for TOC analysis
- Accurate measurement independent of flow rate, pressure or temperature
- Highly reliable, rugged,all-metal design
- No reagents or chemicals required
- No external controller required for analyzer operation
General Specifications
Depth | 4.7 in |
Measurement Range | 1 to 1000 ppb |
On-Line Monitoring | 1 |
Height | 7.6 in |
Width | 16.5 in |
NDIR | 0 |
Power Requirements | Power 100 to 240 VAC, ±10%, 50/60 Hz |
Weight | 10.7 lbs |
Portable | 0 |
Additional Specifications
Measurement Range: 1 to 1000 ppb
Repeatability: ±1% or 1 ppb, whichever is greater
Display Resolution: 1 ppb
Conductivity Accuracy: 2% over full range (uncompensated)
Conductivity Range: 0.05 to 150 ΜS/cm (@ 25 °C)
Conductivity Display Resolution: 0.01 ΜS/cm over full range
Resistivity Range: 0.01 to 18 MW-cm (@ 25 °C)
Resistivity Display Resolution: 0.1 (up to 14.9 MW-cm), 1.0 (from 15-18 MW-cm)
Maximum Input Conductivity: 0.2 ΜS/cm for all waters, 1.0 ΜS/cm for all neutral waters 5 ΜS/cm for water with C02 as the sole conductive species
Conductivity Modes: Temperature compensated to 25 °C or uncompensated
Temperature Accuracy: ±0.5 °C
Minimum Water Quality: 0.2 to 18.2 MW-cm (0.05 to 5 ΜS/cm)
Water Temperature: 5 to 90 °C (41 to 194 °F)
Flow Rate: 60 mL/min to 300 mL/min
Pressure: 10 to 100 psi (60 to 690 kPa)
Installation Category: 11, IEC 1010