TN-4110 On-line TN Analyzer

The TN-4110 is a high performance TN analyzer which uses the combustion / chemiluminiscence detection technique.
Active Questions & AnswersAsk a Question

There are no current Discussions

TOC Analyzers Service ProvidersView All (11)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of TN-4110
  • The TN measurement range is from 0-1 ppm to 0-200 ppm, with the dilution function up to 4000 ppm.
  • Automatic dilution expands the analysis range and allows the reduction in salt, acid, and alkali concentrations to help reduce maintenance and extend catalyst life.
  • A complete line of options designed to accommodate samples with suspended particles, reduce remedial maintenance in on-line TN measurement. 3 type of sample pretreatment units are available.
  • Automatic calibration is incorporated using standard solutions stored on-board. Automatic calibration is possible even when using the on-board dilution function.
  • On-board carrier/combustion gas purification allows use of air supplied from a compressor, thus eliminating expensive and potentially dangerous cylinder gases.
  • Various input/output and alarm function allows a complete control via control station.
General Specifications
On-Line Monitoring1
Power Requirements 100-240 V as ordered; 50-60 Hz
Additional Specifications

Measurement type: Total Nitrogen TN
Measurement principle: Combustion decomposition with chemiluminescence detection.
Measurement range: TN: Variable from 0-1 ppm full scale. 0-4000 ppm is possible with on-board dilution function.
Reproducibility: TN: Within +/-4 % full scale at less than 4 ppm full scale. Within +/-2 % full scale at greater than 4 ppm full scale.
Span stability: Within +/-2 % of full scale/day (with temperature fluctuation within 5 °C)
Zero stability: Within +/-2 % of full scale/week
Measurement cycle: Minimum 4 minutes.
Sample dilution: Dilution with syringe, variable sample dilution factor of 2-20
Auto calibration: 1-2 point auto calibration possible
Multi-stream option: 2 streams with optional sample switching set, Maximum 3 streams with optional flow line switcher
Display: LCD 40 characters x 14 lines
Recorder option: Recorder width is 100 mm, 6 dot recorder
Printer option: Thermal type, 40 columns, chart width with 110 mm.
Analog output: Selectable output of 0-1VDC, 0-16 mA or 4-20 mA (insulated output)
Output: RS-422 / RS-232C(Option)
Alarm: Selectable among upper/lower limit, upper/upper limit and lower/lower limit and instrument trouble alarm
Input signal: Sample measurement start, sample measurement stop, calibration start, alarm reset
Carrier gas: Air or oxygen, supply pressure: 300 KPa, free of dust, oil mist, water droplets
Construction: Indoor wall mounting (optional stand available)