Shimadzu - TOCN-4110
TOCN-4110 On-line TOC/TN Analyzer
Active Questions & AnswersAsk a Question
There are no current Discussions
TOC Analyzers Service ProvidersView All (11)
Documents & Manuals
There are no Documents or Manuals available.
Features of TOCN-4110
- The TOCN-4110 is a compact, unique quality analyzer allowing both TOC analyses by combustion oxidation / infrareddetection method and TN analysis by the combustion / chemiluminiscence detection method in a single unit.
- TOC and TN are simultaneously measured and results obtained in as little as 4 minutes.
- For TOC measurements all known measurements methods (TC,NPOC (standard) TOC=TC-IC (option), TOC=POC+NPOC(option)) are available.
- For TOC a wide measurement range is variable from 0-5ppm full scale to 0-1000 ppm full scale. Measurements up to 20,000 ppm are possible using on-board dilution function. The TN measurement range is from 0-1 ppm to 0-200 ppm, with the dilution function up to 4000 ppm.
- Automatic dilution expands the analysis range and allows the reduction in salt, acid, and alkali concentrations to help reduce maintenance and extend catalyst life.
- A complete line of options designed to accommodate samples with suspended particles, reduce remedial maintenance in on-line TOC/TN measurement. 3 type of sample pretreatment units are available.
- Automatic calibration is incorporated using standard solutions stored on-board. Automatic calibration is possible even when using the on-board dilution function.
- On-board carrier/combustion gas purification allows use of air supplied from a compressor, thus eliminating expensive and potentially dangerous cylinder gases.
- Various input/output and alarm function allows a complete control via control station.
General Specifications
On-Line Monitoring | 1 |
Power Requirements | 100-240 V as ordered; 50-60 Hz |
Additional Specifications
Measurement type: NPOC(acidify/sparge removal of IC) and TC (standard). NPOC, TOC (TC-IC) (option). NPOC,TOC (TC-IC and POC + NPOC) (option) Total Nitrogen TN
Measurement principle: TOC: Catalytically aided combustion with NDIR detection,
TN: combustion decomposition with chemiluminescence detection.
Measurement range: TOC: Variable from 0-5 ppm full scale to 0-1000 ppm full scale (0-20,000 ppm is possible with on-board dilution function).
TN: Variable from 0-1 ppm full scale. 0-4000 ppm is possible with on-board dilution function.
Reproducibility: TOC: Within +/-2 % full scale. TN: Within +/-4 % full scale at less than 4 ppm full scale. Within +/-2 % full scale at greater than 4 ppm full scale.
Span stability: Within +/-2 % of full scale/day (with temperature fluctuation within 5 °C)
Zero stability: Within +/-2 % of full scale/week
Measurement cycle: Minimum 4 minutes.
Sample dilution: Dilution with syringe, variable sample dilution factor of 2-20
Auto calibration: 1-2 point auto calibration possible
Multi-stream option: 2 streams with optional sample switching set, Maximum 3 streams with optional flow line switcher
Display: LCD 40 characters x 14 lines
Recorder option: Recorder width is 100 mm, 6 dot recorder
Printer option: Thermal type, 40 columns, chart width with 110 mm.
Analog output: Selectable output of 0-1VDC, 0-16 mA or 4-20 mA (insulated output)
Output: RS-422 / RS-232C(Option)
Alarm: Selectable among upper/lower limit, upper/upper limit and lower/lower limit and instrument trouble alarm
Input signal: Sample measurement start, sample measurement stop, calibration start, alarm reset
Carrier gas: Air or oxygen, supply pressure: 300 KPa, free of dust, oil mist, water droplets
Construction: Indoor wall mounting (optional stand available)