Shimadzu - UV-3600

Manufactured by  Shimadzu
Follow this Equipment

UV-VIS Spectrophotometer High Sensitivity, High Resolution, Ultra-low Stray Light

The UV-3600 UV-VIS-NIR Spectrophotometer as a new additional model in our product lineup of high-end UV-VIS-NIR spectrophotometers.

The UV-3600, a successor model to the UV-3150/3101, is designed for the measurement of liquid samples, while the SolidSpec-3700/3700DUV is mainly designed for the measurement of solid samples.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked byfram

UV/VIS Spectrophotometers Service ProvidersView All (32)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of UV-3600
  • High Sensitivity - 3 Detector System Equipped with 3 detectors: a PMT detector (photomultiplier tube) for the ultraviolet and visible regions, and InGaAs and PbS detectors for near infrared region. The InGaAs detector bridges the gap between the PMT - PbS switching wavelength where sensitivity is typically low to ensure high sensitivity over the entire measurement wavelength range. With <0.00003 Abs noise at 1500 nm, the world’s lowest noise level is achieved.
  • High Resolution - Wide Measurement Range & Ultra-low Stray Light With its high performance double monochromator, ultra-low stray light (0.00005% or less at 340 nm) is achieved at high resolution (highest resolution 0.1 nm). With a measurement wavelength range of 185 - 3300 nm, measurement is possible over a wide range including the ultraviolet, visible and near infrared regions. This allows spectroscopic analysis in a wide variety of fields.
  • Abundant Accessories for a Wide Variety of Applications The multipurpose large-sample compartment and integrating sphere attachment enable measurement of solid samples, and with the guaranteed precision of the ASR Series (Absolute Specular Reflectance) attachments, high accuracy absolute reflectance measurements are possible. Various cell holders like the constant-temperature cell holder and ultra-micro cell holder can also be mounted to enable a wide range of measurement possibilities.
General Specifications
Bandwidth8 steps in ultraviolet/visible region 0.1, 0.2, 0.5, 1, 2, 3, 5, 8 nm; 10 steps in near-infrared region 0.2, 0.5, 1, 2, 3, 5, 8, 12, 20, 32 nm
Depth 660 mm
Height275 mm
Width 1020 mm
Photometric Range-6 to 6 Abs
Wavelength Range185 to 3300 nm
OpticsDouble Beam
Photodetector Photomultiplier
Photometric Noise0.00005 Abs or less (500 nm), 0.00008 Abs or less (900 nm), 0.00003 Abs or less (1500 nm). Slit width 2nm, RMS value at 1 sec. response
Additional Specifications

Resolution: 0.1 nm
Sampling Pitch: 0.01 - 5 nm
Wavelength accuracy: UV/VIS region: ±0.2 nm NIR region: ±0. 8 nm
Wavelength repeatability: UV/VIS region: less than ±0.08 nm NIR region: less than ±0.32 nm
Stray light:
< 0.00008% (220 nm, NaI)
< 0.00005% (340 nm, NaNO2)
< 0.0005% (1420 nm, H2O)
< 0.005% (2365 nm, CHCl3)
Photometric System: Double beam

Looking for New or Used Equipment?

Shop on LabX

Shop on Labx.com