Shimadzu - RF-5301PC
Compared to absorbance methods, fluorescence sensitivity is tens to thousands times better - this means that you can...
Fluorescence can be used also to identify a specific molecule in a complex background. When the compound of interest does not exhibit natural fluorescence, functional group-specific probes may be used to label the compound and assist your research. The synchronous scanning mode allows mixtures of fluorochromes to be analyzed. The personal computer directly controls the instrument for data acquisition and processing. The Windows friendly operating environment allows you to perform measurement, data processing, editing and recording in one continuous operation with a click of the mouse. Using the Copy Graph function, measurement data or spectra may be easily transferred to word processing or spreadsheet software for preparation of documents or additional calculations.
The essence of fluorescence analysis is sensitivity. The high throughput optical system in the RF-5301PC employs a blazed holographic grating, photomultiplier and digital signal processing to provide the highest level S/N ratio attainable.
Active Questions & AnswersAsk a Question
There are no current Discussions
Need Equipment Support?
Documents & Manuals
There are no Documents or Manuals available.
Features of RF-5301PC
- High Speed Scanning
- High Resolution and Extended Range
- Built-In Performance Checks
General Specifications
Depth | 530 mm |
Height | 270 mm |
Width | 667 mm |
Power Requirements | 100, 120, 220, 240V; 50/60Hz |
Peak Power | 400 VA |
Weight | 43 kg |
Additional Specifications
Light Source: 150W Xenon lamp. Ozone resolving type lamp housing.
Excitation and emission monochromators: Concave, blazed holographic grating, F/2.5, 1300 grooves/mm.
Wavelength scale: 220-900nm.
Measuring wavelength range: 220-750nm and 0 order as standard. 220-900nm with the optional R928 photomultiplier.
Spectral bandwith: 6-step selection of 1.5, 3, 5, 10, 15 and 20nm. (6nm bandwith is available for half sample height on the excitation side only.)
Wavelength accuracy: ±1.5nm.
Light source compensation: Dynode feedback system with monochromatic light monitoring.
Sensitivity: The S/N ratio is 150 or higher for the Raman line of distilled water (350nm excitation wavelength, 5nm spectral bandwidth, and 2 second response for 98% of the full scale).
Wavelength scanning: 7-step selection of Survey (about 5500nm/min), Super (about 3000nm/min), Very Fast, Fast, Medium, Slow and Very Slow.
Wavelength slewing speed: About 20,000nm/min.
Response: 8-step selection of 0.02, 0.03, 0.1, 0.25, 0.5, 2, 4, and 8 seconds for 98% of the full scale.
Sensitivity selection: 2-steps of HIGH and LOW. (The sensitivity at HIGH is about 50 times that of LOW.)
Interface: RS-232C interface, interface for autosampler, and interface for sipper unit.
Operational temperature range: 15-35°C
Operational humidity range: 40-80% (Below 70% with temperature higher than 30°C).