Shimadzu - AOC Series
The AOC-20i auto injector is the perfect solution for automated analyses.
Active Questions & AnswersAsk a Question
GC Autosamplers Service ProvidersView All (14)
Documents & Manuals
There are no Documents or Manuals available.
Features of AOC Series
Simplified Maintenance The AOC-20 is highly reliable and capable of high sample throughput, yet it is lightweight and easy to remove when routine injection port maintenance is required. The enhanced needle guide virtually eliminates bent syringe needles; syringe changes are easy to accomplish through the front door of the unit.
Reliable Injection Performance Possible injection modes include traditional, solvent flush, solvent flush with second solvent, and large volume injection. The AOC-20 series can perform the functions of direct liquid injection, on-column injection, ambient headspace and large volume injection. These capabilities are the most available from a single autosampler. The high precision of the injector has been demonstrated by reproducibility studies (Grob test mixture) that yielded C.V. values of less than 0.3%!
Complete Control The AOC-20 offers control through fiber optic or RS-232C cable. Specify and save parameters from PC-based systems or from the Shimadzu C-R8A Chromatopac data processor..
General Specifications
There are no General Specifications available.
Series Modals
Shimadzu - AOC-20i
[+] View Model SpecificationsDepth | 78 mm |
Height | 380 mm |
Width | 126 mm |
Power Requirements | AC100/115/220V±10% 50/60HZ |
Number of Samples | 6 vials |
Weight | 2.5 kg |
Additional Model Specifications | Sample Injection Method: Liquid sample injection via special microsyringe Sample Volume: 0.1 ~ 8.0 µl, 0.1µl steps (using 10ul syringe) 0.5 ~ 40 µl, 0.5 µl steps (using 50ul syringe) 2.5 ~ 200 µl, 2.5 µl steps (using 250ul syringe) Sample Vials: Glass construction, 1.5ml, 4ml, screw top, PTFE-coated septum Rinse Solvent Vials: Glass construction, 4ml, screw top, PTFE-coated septum Number of Sample Injections: 1 ~ 99 injections per sample Syringe Speed: 2 Modes: Fast and Slow Plunger Speed: 3 Modes: Fast, Medium and Slow Wait Time: 0 ~ 99 sec following sample aspiration (in 0.1 sec steps) Type of Sample Injection: 3 Modes: Traditional, Solvent Flush and Solvent Flush with a Second Solvent Injection Volume Linearity: ±0.5% (injection volume 1 ~ 5 µl, sample n-C12 Cross Contamination: Less than 10-4 (as determins with a 1% diphenyl in hexane using 4 solvent rinses) Priority Sample: A priority sample injection can be injected during a sample sequence, the the sequence can be resumed Dual Injection System: Available on GC-17A platform only External Control: Includes optional optical link or RS232C interface |
Shimadzu - AOC-20s
[+] View Model SpecificationsHeight | 135 mm |
Width | 320 mm |
Power Requirements | AC100/115/220V±10% 50/60HZ |
Number of Samples | 1.5ml sample vials - 150 vials |
Weight | 2.4 kg |
Additional Model Specifications | Sample Injection Method: Liquid sample injection via special microsyringe Sample Volume: 0.1 ~ 8.0 µl, 0.1µl steps (using 10ul syringe) 0.5 ~ 40 µl, 0.5 µl steps (using 50ul syringe) 2.5 ~ 200 µl, 2.5 µl steps (using 250ul syringe) Sample Vials: Glass construction, 1.5ml, 4ml, screw top, PTFE-coated septum Rinse Solvent Vials: Glass construction, 4ml, screw top, PTFE-coated septum Number of Sample Injections: 1 ~ 99 injections per sample Syringe Speed: 2 Modes: Fast and Slow Plunger Speed: 3 Modes: Fast, Medium and Slow Wait Time: 0 ~ 99 sec following sample aspiration (in 0.1 sec steps) Type of Sample Injection: 3 Modes: Traditional, Solvent Flush and Solvent Flush with a Second Solvent Injection Volume Linearity: ±0.5% (injection volume 1 ~ 5 µl, sample n-C12 Cross Contamination: Less than 10-4 (as determins with a 1% diphenyl in hexane using 4 solvent rinses) Priority Sample: A priority sample injection can be injected during a sample sequence, the the sequence can be resumed Dual Injection System: One set of AOC-20s can feed sample vials to two sets of AOC-20i External Control: Includes optional optical link or RS232C interface Sample Cooling/Heating: 0 ~ 60°C, with optional cooling rack connected to a general laboratory circulating water bath |