BINDER - MK series
Manufactured by BINDER
Environmental simulation chamber for complex temperature profiles
The BINDER MK series chamber is well-suited for all heating and cooling testing between -40 °C and 180 °C. The APT.line™ pre-heating chamber technology uniquely simulates a natural environment. For cyclical temperature testing, it is a smart alternative to complex individual solutions.
Active Questions & AnswersAsk a Question
There are no current Discussions
Environmental Chambers Service ProvidersView All (20)
Documents & ManualsView All Documents
Features of MK series
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- User-friendly LCD screen -Easy-to-read menu guide -Integrated electronic chart recorder -Variety of options for the graphic display of process parameters -Real-time clock
- Programmable condensation protection for test material
- Powerful adjustable fan
- Adjustable ramp function via program editor
- Access port Ø 80 mm, top
- Independent adjustable temperature safety device class 2 (DIN 12880), with visual and audible temperature alarm
- Heated viewing window with LED interior lighting
- Environmentally friendly refrigerant R 404a
- RS 422 interface for communication software APT-COM™ DataControlSystem
- 1 shelf, stainless steel
- BINDER test certificate
General Specifications
There are no General Specifications available.
Series Modals
BINDER - MK 115
[+] View Model SpecificationsDepth | 855 mm |
Height | 1718 mm |
Width | 995 mm |
Weight | 260 kg |
Additional Model Specifications | Interior dimensions Width (mm): 600 Height (mm): 480 Depth (mm): 400 Interior volume (l): 115 Shelves (number standard/max.): 1 / 4 Load per shelf (kg): 30 Permitted total load (kg): 60 Temperature data Temperature range (°C) 1): -40 - 180 Temperature variation (± K): 0,1 - 2,0 Temperature fluctuation (± K): 0,1 - 0,5 Mean warm-up rate acc. to IEC 60068-3-5 (°C/min.): 5,5 Mean cooling rate acc. to IEC 60068-3-5 (°C/min.): 5,2 Warm-up time from -40 °C to 180 °C (min.): 45 Cooling down time from 180 °C to -40 °C (min.): 90 Heat compensation, max. (W): 2000 Electrical data IP protection class acc. to EN 50529: IP 20 Voltage 50 Hz (V): 400 3N~ Nominal power (kW): 3,0 Energy consumption at 20 °C (W) 1): 600 Noise level (approx. dB(A)): 62 |
BINDER - MK 240
[+] View Model SpecificationsDepth | 946 mm |
Height | 1713 mm |
Width | 1130 mm |
Weight | 360 kg |
Additional Model Specifications | Interior dimensions Width (mm): 735 Height (mm): 700 Depth (mm): 443 Interior volume (l): 228 Racks (number standard/max.): 1/ 6 Load per shelf (kg): 30 Permitted total load (kg): 70 Temperature data Temperature range (°C) 1): -40 - 180 Temperature variation (± K): 0,1 - 1,2 Temperature fluctuation (± K): 0,1 - 0,5 Recovery time after door was open for 30 sec.: at -10 °C (min.): 8 at 70 °C (min.): 2 at 150 °C (min.): 7 Mean warm-up rate acc. to IEC 60068-3-5 (°C/min.): 5 Mean cooling rate acc. to IEC 60068-3-5 (°C/min.): 4.5 Warm-up time from -40 °C to 180 °C (min.): 50 Cooling down time from 180 °C to -40 °C (min.): 110 Heat compensation, max. (W): 2000 Electrical data IP protection class acc. to EN 50529: IP 20 Voltage 50 / 60 Hz (V): 400 3N~ Nominal power (kW): 4,2 Energy consumption at 20 °C (W) 1): 1300 Noise level (approx. dB(A)): 62 |
BINDER - MK 53
[+] View Model SpecificationsDepth | 794 mm |
Height | 1242 mm |
Width | 740 mm |
Weight | 150 kg |
Additional Model Specifications | Interior dimensions Width (mm): 402 Height (mm): 402 Depth (mm): 330 Interior volume (l): 53 Shelves (number standard/max.): 1 / 5 Load per shelf (kg): 15 Permitted total load (kg): 40 Temperature data Temperature range (°C): -40 - 180 Temperature variation (± K): 0,4 - 2,0 Temperature fluctuation (± K): 0,1 - 0,5 Recovery time after door was open for 30 sec.: at -10 °C (min.): 5 at 70 °C (min.): 1 at 150 °C (min.): 5 Mean warm-up rate acc. to factory standard (°C/min.) -40 °C to 180 °C: 4,6 Mean cooling rate acc. factory standard (°C/min.) 180 °C to -40 °C: 4,1 Warm-up time from -40 °C to 180 °C (min.): 47 Cooling down time from 180 °C to -40 °C (min.): 93 Heat compensation, max. (W): 500 Electrical data IP protection class acc. to EN 50529: IP 20 Voltage (± 10%) 50 Hz (V): 230 1N~ Nominal power (kW): 2,6 Energy consumption at 20 °C (kW) 1): 1,02 Noise level (approx. dB(A)): 59 |
BINDER - MK 720
[+] View Model SpecificationsDepth | 1175 mm |
Height | 2005 mm |
Width | 1615 mm |
Weight | 570 kg |
Additional Model Specifications | Interior dimensions Width (mm): 1200 Height (mm):1020 Depth (mm): 600 Interior volume (l): 734 Shelves (number standard/max.): 1 / 11 Load per shelf (kg): 40 Permitted total load (kg): 160 Temperature data Temperature range (°C): -40 - 180 Temperature variation (± K): 0,3 - 2,0 Temperature fluctuation (± K): 0,1 - 0,5 Mean warm-up rate acc. to IEC 60068-3-5 (°C/min.): 4 Mean cooling rate acc. to IEC 60068-3-5 (°C/min.): 4,5 Warm-up time from -40 °C to 180 °C (min.): 58 Cooling down time from 180 °C to -40 °C (min.): 75 Heat compensation (W): 6500 Electrical data Voltage (± 10%) 50 Hz (V): 400 3N~ Nominal power (kW): 7,2 Energy consumption at 20 °C (kW) 1): 1,9 Noise level (approx. dB(A)): 65 |