Agilent Technologies - 1260 Infinity Thermostatted Column Compartment
Manufactured by Agilent Technologies
Convenient, stable column cooling and heating
The Agilent 1260 Infinity Thermostatted Column Compartment provides stable column cooling and heating up to 80 °C for standard applications. It fits neatly in a single stack of 1260 Infinity modules for lowest bench space requirements. An optional 2-position/6-port switching valve is available for 600 bar operation. For more sophisticated applications the Agilent 1290 Infinity Thermostatted Column Compartment is available, which fits seamlessly into any 1260 Infinity system.
Active Questions & AnswersAsk a Question
There are no current Discussions
Column Ovens Service ProvidersView All (16)
Documents & ManualsView All Documents
Features of 1260 Infinity Thermostatted Column Compartment
- Peltier cooling and heating for maximum temperature stability and application flexibility.
- Spacious column racks hold up to three full-length, 30 cm columns.
- Optional 2-position/6-port switching valve for 600 bar operation.
- Simultaneous column identification of two columns for faster operation and reduced laboratory costs.
- Independent heat exchangers with settable values facilitate two different temperature zones.
- Low internal volume contribution of just 9 µL for low dispersion.
General Specifications
There are no General Specifications available.
Additional Specifications
Temperature range:
10 degrees below ambient to 80 °C
Temperature stability:
± 0.15 °C
Temperature accuracy:
± 0.8 °C with calibration ± 0.5 °C
Column capacity:
Three 30 cm columns
Heat-up/cool-down time:
5 min from ambient to 40 °C
10 min from 40 °C to 20 °C
Internal volume:
3 µL left heat exchanger
6 µL right heat exchanger
GLP:
Column identification module for GLP documentation of column type