HORIBA - XGT-1000WR

Manufactured by  HORIBA
Follow this Equipment

X-ray Fluorescence Analyzer

XGT-1000WR are specially designed for WEEE/RoHS, ELV and Chinese RoHS for the high sensitivity measurement of 5 elements (Pb/Cd/Cr/Hg/Br).

Simply place the sample by hand directly into the large sample chamber (460 x 360 x 150 mm), and then easily position the measurement location by hand while verifying the location with the CCD camera image. Directly measures parts and products without sample preparation. The XGT-1000WR has been designed to simplify inspections such as analyses performed on manufacturing lines.

These instruments are equipped with a standard 1.2 mm-diameter high-intensity X-ray guide tube (XGT:HORIBA original) and with fixed primary filter for high sensitivity for the 5WR elements. XGT provides shorter measurement time and greater precision than methods using an X-ray collimator.

When the guide tube is switched to the optional 100 µm X-ray guide tube with possibility to use the vacuum pump to determine light elements, it can also be used as micro-area element analyzers and offer sensitivity to all elements from Sodium (Na) to Uranium (U).
Active Questions & AnswersAsk a Question

Recent Questions & Answers

Need Equipment Support?

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of XGT-1000WR
  • HORIBA original X-ray guide tube - High-precision and high speed analysis
  • Easy sample setting with CCD camera
  • No need vacuum in the sample chamber
  • Compliant with WEEE/RoHS directives
  • Compatible with a wide range of sample size from small parts to large products
General Specifications
Depth 750 mm Analyzing unit
Height500 mm Analyzing unit
Width 610 mm Analyzing unit
Power Requirements AC100, 120, 220, 240V 50/60Hz
Weight200 kg
Additional Specifications

Outer dimensions
Signal Processing unit: 220(W)x500(D)x480(H)mm
Sample chamber: 460mm×360mm×150mm
Principle:
Energy dispersive X-ray fluorescence analyzer
Element:
Si to U and Na to U (Optional)
Sample:
Plastic, metal, paper, liquid such as paint and ink etc
Sample dimensions:
Max 460mm×360mm (Height 150mm)
Sample chamber atmosphere:
Atmosphere
XGT Element Spatial Resolution:
1,2mm and additional 100µm (Optional)
X-ray tube:
50kV/1mA, Rh target
Vacuum probe:
Optional
Primary X-ray Filter:
Ø1,2mm: 5WR elements simultaneous filter (Cd/Pb/Cr/Hg/Br)
Ø1,2mm (Optional): Multifilter with 4 specific filters (Cd/Cr,Cl/5WR elements/No filter)
Ø100µm:Filter less
Secondary X-ray filter:
None
Detector:
High purity Si detector

Looking for

HORIBA - XGT-1000WR?

Shop on LabX.com