Varian - 940-LC Analytical to Prep HPLC
Scale Up to Superior Performance and Reliability
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Updated byLabWrenchEditor
Preparative HPLC Service ProvidersView All (16)
Documents & ManualsView All Documents
Features of 940-LC Analytical to Prep HPLC
- High performance analytical HPLC
- Fast, easy scale-up to preparative HPLC
- Flexible configurations
- Unmatched concentration range
- Versatile detection
- Superb, practical fraction collection
- Simple, productivity-boosting software
General Specifications
Depth | 725 mm |
Height | 665 mm |
Optional Flow Rate | 0.2 to 200 ml/min |
Width | 750 mm |
Power Requirements | 90/120 V AC or 220/250 V AC, 50/60 Hz |
Peak Power | 1200 VA |
Standard Flow Rate | 0.025 to 25 ml/min |
Weight | 137 kg |
Additional Specifications
High pressure gradient solvent delivery system
Type: 2 x reciprocating single piston, rapid refill design
Maximum pressure: With Extended Pressure Option
32 MPa (4600 psi, 320 bar) from 0.025 to 25 mL/min
14 MPa (2000 psi, 140 bar) to 50 mL/min
8 MPa (1200 psi, 80 bar) to 100 mL/min
41 MPa (6000 psi, 410 bar) from 0.025 to 25 mL/min
31.5 MPa (4000 psi, 315 bar) to 100 mL/min
24.4 MPa (3500 psi, 244 bar) to 200 mL/min
Number of solvents: Maximum of 4
Flow rate range:
0.05 to 50 mL/min with 50 mL heads
0.1 to 100 mL/min with 100 mL heads
Flow accuracy: 1% of selected flow or 0.05% of maximum flow, whichever is larger
Flow reproducibility: 1% of selected flow or 0.05% of maximum flow, whichever is larger
Autosampler
Maximum number of samples:
128 x 4.5 mL HPLC vials
200 x 2 mL HPLC vials
3 x 96-well wellplates
Autosampler syringe volume: 5 mL
Injection volume range: 0.005 to 4.5 mL
Injection repeatibility: 0.5% RSD (using a 500 ΜL injection by a 5 mL syringe)
Carryover: <0.01%
Temperature-controlled autosampler (optional): Settable temperature range 4 to 35 °C (in increments of 1 °C)
Cooling system: Peltier effect-based thermoelectronic cooling type
UV detector
Light source: Deuterium
Wavelength range: 190 to 700 nm (standard); 190 to 900 nm (optional)
Spectral bandpass: 6 nm
Wavelength accuracy: ±1 nm at 260 nm
Wavelength precision: ±0.1 nm
Noise: ±5x10-6 AU
Drift: <1x10-3 AU/h
Data acquisition rate: 20 Hz single wavelength mode
Maximum pressure rating: Flow cell rated to 7 MPa (1000 psi, 70 bar)
Flow cells:
4 x 0.15 mm dual pathlength super prep (standard for preparative system)
4 x 0.25 mm dual pathlength scale-up (standard for semi-preparative system)
9 x 1 mm semi-preparative dual pathlength
Diode dual array detector
Light source: Deuterium for UV and quartz halogen for visible
Wavelength range: 190 to 950 nm
Wavelength accuracy: ±1 nm verified with 486 nm D2 line
Spectral resolution: <1 nm per diode
Digital resolution: 1 nm
Noise: ±10x10-6 AU
Drift: <1x10-3 AU/h
Data acquisition rate: 10 Hz
Maximum pressure rating: Flow cell rated to 7 MPa (1000 psi, 70 bar)
Flow cells:
4 x 0.15 mm dual pathlength super prep (standard for preparative system)
4 x 0.25 mm dual pathlength scale-up (standard for semi-preparative system)
9 x 1 mm semi-preparative dual pathlength
ELS detector (external, optional)
Detector Blue LED 480 nm. Class 1 LED product.
Temperature range:
Evaporator: Ambient to 120 °C (10 to 80 °C, cooled ELS detector)
Nebulizer: Ambient to 90 °C
Nebulizer gas:
Typical inlet pressure: 414 to 689 kPa (60 to 100 psi, 4 to 7 bar)
Maximum inlet pressure: 689 kPa (100 psi, 7 bar)
Gas flow rate: Up to 3.25 SLM
Eluent flow rate: Up to 5 mL/min
Digital output: 24 bit (10 Hz)
Communications:
Serial I/O (RS232), contact closure, TTL
Power requirements: 90/120 V AC or 220/250 V AC, 50/60 Hz, VA 1200
Dimensions:(W x D x H) 200 x 450 x 415 mm
Weight: 11 kg (13 kg, cooled ELS detector)
Scale-Up Module
High-pressure valves:
Port size: Accepts 10-32 male-threaded fittings
Flow passage diameters: Stator 0.3 mm and 0.25 mm for vertical port
Rotor seal: 0.25 mm
Port-to-port volume: 445 nL and 586 nL with vertical port
Maximum pressure: 41 MPa (6000 psi, 410 bar)
Actuation time: 100 ms
Low-pressure valves:
Orifice diameter: 0.81 mm
Internal volume: 45 ΜL
Ports: 1/4-28 flat bottom
Operating pressure: Vacuum - 70 kPa (10 psi, 0.7 bar) (normally open/normally closed)
Vacuum - 172 kPa (25 psi, 1.7 bar) (common)
Wetted parts: Perfluoroelastomer and PEEK™
Fraction collector
Maximum sample racks: 3
Typical number of containers: 180
Maximum tube height: 150 mm
Rack options:
Type 21 - 21 x 30 mm OD containers
Type 24 - 24 x 25 mm OD containers
Type 40 - 40 x 20 mm OD containers
Type 60 - 60 x 16 mm OD containers
Type 90 - 90 x 13 mm OD containers
Type 128 - 128 x 15 mm OD containers
Type 200 - 200 x 12 mm OD containers
Dimensions: (W x D x H) 400 x 600 x 480 mm
Weight: 18 kg
Divert valve:
Orifice diameter: 0.81 mm
Internal volume: 45 ΜL
Ports: 1/4-28 flat bottom
Operating pressure: Vacuum - 70 kPa (10 psi, 0.7 bar) (normally open/normally closed)
Vacuum - 172 kPa (25 psi, 1.7 bar) (common)
Wetted parts: Perfluoroelastomer and PEEK
Manual injector (optional)
Port size: Accepts 10-32 male-threaded fittings
Flow passage diameters: Stator 0.3 mm and 0.25 mm for vertical port
Rotor seal: 0.25 mm
Port-to-port volume 445 nL and 586 nL with vertical port
Maximum pressure 41 MPa (6000 psi, 410 bar)
Actuation time 100 ms