Features of LC-2010HT
Large sample capacity The LC-2010HT can handle up to 350 1mL vials or 210 1.5mL vials1 . When microtiter plate racks are used, up to 4 microtiter plates can be accommodated for continuous analysis of max. 1,536 samples.2 Deep Well plates can also be used. Two separate sample racks are used so that you can load samples in a rack while analyses are performed on samples from the other rack.
High-speed injection Injection time per sample has been greatly reduced. Just 15 seconds* are required from start to finish for unheard of high-speed sample injection.
High-speed analysis Seven analyses are completed in 15 minutes thanks to high-speed injection and an ultra-fast column! This enables major improvements in analysis throughput in the field of drug discovery.
Database Multiple data are managed in one block with optional CLASS-Agent Manager*. Required data can be easily retrieved from a mass of data as analysis data is automatically loaded into the database.
General Specifications
Depth | 500 mm |
Detector | UV |
Height | 605 mm |
Width | 430 mm |
Power Requirements | AC 100V-240V, 50/60Hz |
Peak Power | 700 VA |
Weight | 52 kg |
Additional Specifications
Pump
Pump Type: Serial dual plunger, micro volume (10µL on primary side, 5µL on secondary)
Flow Rate: 0.001-5mL/min [1.0-35MPa / 145-5076psi]
Flow Rate Accuracy: ±1% or ±2µL/min, whichever is larger (1mL/min, 3-10.0MPa / 435-1450psi, water, ambient temperature at 20°C, except in gradient solvent delivery),±2% or ±2µL/min, whichever is larger (except above condition and in gradient solvent delivery)
Flow Rate Precision: ±0.075% RSD or 0.02minSD, whichever is larger (0.05-2mL/min, 1.0-3.5MPa / 145-5076psi, methanol, ambient temperature at 20°C
Pressure Display Accuracy ±2% or ±0.5MPa, whichever is larger
Degasser
Type: Membrane on-line degasser, 5 lines (4x mobile phase, 1x autosampler rinsing liquid)
Volume: 4mL/min for mobile phases, 2mL/line for autosampler rinsing liquid
Autosampler
Injection Method: Full volume injection, variable injection possible (no sample loss)
Injection Volume: 0.1 - 100µL (standard), 1-2000µL (optional)
(0.1-0.9µL in 0.1µL step, 1-2000µL in 1 µL step)
Sample Capacity:
1 mL vials: 350
1.5mL vials: 140
4mL vials: 100
Microtiter Plates: Max. 4 plates (up to 384 samples with 96 well or 1536 samples with 384 well plates)
Deep Well Plates: Max. 4 plates (up to 384 samples with 96 well plates)
Injection Volume Repeatability: RSD <0.3% (10µL injection)
Carryover: 0.01% (under conditions determined by Shimadzu)
Injection Volume Accuracy: ±1% (50µL, n=6)
Repetition Injection Frequency: 1-99/sample
Analytical Run Time Setting: 0.01-9999.9 min (0.01 min step)
Injection Condition Steps: Max. 202 steps
Operating Pressure: 35.0MPa maximum
Rinsing Liquid Degassing: Standard
Column Oven
Type: Block heating with preheating of the mobile phase
Temperature Setting Range: 4-60°C, 1°C step
Temperature Control Precision: ±0.1°C
Temperature Control Range: (Ambient temperature -15°C) - 60°C
Columns accommodated: 2 x 250mm columns
Preheater volume: 4µL (for Semi-micro), and 9µL, 2 channels
UV Detector
Light Source: Deterium lamp, Low pressure mercury lamp (for wavelength accuracy check)
Wavelength Range: 190-600nm
Bandwidth: 8nm
Wavelength Accuracy: ±1nm
Wavelength Reproducibility: ±0.1nm
Cell: 10mm, 8µL
Cell Temperature Control: Standard (Setting temperature: High, Low, and OFF, 3 steps)
Drift: 1 x 10-4AU/h (250nm, air in cell, constant room temperature)*
Noise Level: ±0.25 x 10-5 AU (2 sec. time constant, 250nm, air in cell)*
Linearity: ~ 2.5AU
Simultaneous Dual Wavelenth Measurement: Selectable, two wavelengths in 190 - 370nm or 371 - 600nm
Signal Output: 2 channels
Automation Function: Auto start-up, Auto shut-down, Baseline stability autojudgement, Auto validation (Wavelength, Lamp Energy, Solvent delivery plusation, Column Temperature, Absorbance, Baseline drift, Baseline noise, Pressure Limit, Gradient accuracy)
Auto purge, Cell auto recognition, Column management (option)
Memory check, Max/Min pressure, Oven max. temperature, Lamp current, Motor Revolution monitor, Leak Sensor, Lamp Cover Sensor, Lamp housing max. temperature
Application Solvent Organic solvents (except fluorine organic solvent, e.g. HFIPA), Aqueos solution and mixture of diluted acid or base (except halo-acids, e.g. hydrochloric acid)
pH Range: 1-13 standard
Liquid Contact Surface Materials: SUS316, Ruby, Sapphire, Teflon®, Hasteloy C, PEEK, Trifluoroethylene, PFA Perfluoroelastomer, Quartz, Ceramic
Operating Ambient Temperature Range, Humidity: 4-35°C; max. 80% humidity
Power Requirements
*Condition according to ASTM E1657-96