Edwards - XDS10 Scroll Vacuum Pump

Manufactured by  Edwards
Follow this Equipment

XDS10 3-phase 200/380/415/460 V set to high volts

The XDS pumps are an innovative design which use a bearing shield to isolate the vacuum environment from all forms of lubricant, not only making it totally dry, but hermetically sealed. The shield also protects the bearing from any process vapors.
Adjustable gas ballast allows vapor to be handled and opens up the range of applications to many that were previously unsuited to scroll pumps. The XDS pumps are designed for pumping condensable vapors and use materials selected for a wide range of laboratory wet chemistry applications.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Updated byhoward2012

Vacuum Pumps Service ProvidersView All (13)

Request Support
Features of XDS10 Scroll Vacuum Pump

Bearing shield

  • Totally clean and dry vacuum
  • No bearings or grease in the vacuum space
  • No shaft seals, no bearing purges
  • Simple single sided scroll design
  • Tip seal change in 20 minutes
  • No special tools required

Selectable gas ballast position

  • No permanent gas ballast leak
  • Large water vapor handling capacity

All pumps externally approved to EN61010, CSA C22.2 1010, UL 61010 standards

  • Ease of incorporating pump into approved systems, safe product

Manufactured under

  • ISO 9001 standard
  • High quality product

Manufactured under ISO 14001 standard

  • Environmentally friendly processes
General Specifications
Vacuum Pump TypeScroll Pump
Weight24.5 kg
Additional Specifications

Displacement
50Hz: 11.3 m³h^-1 / 6.7 ft³min-1
60Hz: 13.4 m³h^-1 / 7.9 ft³min-1

Peak pumping speed
50Hz: 9.3 m³h^-1 / 5.5 ft³min-1
60Hz: 11.1 m³h^-1 / 6.5 ft³min-1

Ultimate vacuum (Total pressure)
50Hz: 7 x 10^-2 mbar / 5.3 x 10^-2 Torr
60Hz: 6 x 10^-2 mbar / 4.5 x 10^-2 Torr

Max inlet pressure for water vapour: 50 mbar / 38 Torr
Max water vapour pumping rate – GB I: 70 gh^-1
Max water vapour pumping rate – GB II: 200 gh^-1
Max allowed outlet pressure: 1 bar gauge
Max allowed inlet and gas ballast pressure: 0.5 bar gauge
Motor power 1-ph and 3-ph: 0.3kW

Nominal rotation speed
50/60 Hz: 1500 / 1800 rpm
Water vapour tolerance: 35 mbar / 23 Torr
Water vapour capacity – GB I: 70 gh^-1
Water vapour capacity – GB II: 200 gh^-1
Motor power 1-ph and 3-ph 300 W
Inlet flange: NW25
Exhaust flange: NW25
Noise level: 55 dBA @ 50 Hz
Vibration: <4.5 mms^-1 (rms)
Leak tightness (static): <1 x 10^-6 mbar ls^-1
Operating temperature range: 10 to 40 °C / 50 to 104 °F

Looking for

Edwards - XDS10 Scroll Vacuum Pump?

Shop on LabX.com