BINDER - MKF series
Manufactured by BINDER
Environmental simulation chamber for complex alternating climate profiles
Environmental simulation chamber for complex alternating climate profiles
A BINDER MKF series chamber is ideally suited for all testing based on current temperature and climate testing methods according to DIN and IEC standards. Their ease of operation and comprehensive standard equipment ensures something special in use.
Areas of application:
A BINDER MKF series chamber is ideally suited for all testing based on current temperature and climate testing methods according to DIN and IEC standards. Their ease of operation and comprehensive standard equipment ensures something special in use.
- State-of-the-art reliability
- User-friendly chamber interior
- Comprehensive standard equipment
Areas of application:
- Automotive
- Electronics / Semiconductor Industry
- Plastics Industry
Active Questions & AnswersAsk a Question
There are no current Discussions
Environmental Chambers Service ProvidersView All (20)
Documents & ManualsView All Documents
Features of MKF series
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range without humidity -40 °C to 180 °C (-40 °F to 356 °F), with humidity 10 °C to 95 °C (50 °F to 203 °F)
- Humidity range 10 % RH to 98 % RH
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- User friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real time clock
- Electronically controlled humidification and dehumidification system with capacitive humidity sensor and pressurized steam humidification
- Integrated water storage tank
- Heated viewing window with interior illumination
- Programable condensation protection for test material
- Adjustable ramp functions via program editor
- Environmental friendly refrigerant R 404a
- 230 V power socket on the right-side operating panel
- Temperature safety device, Class 2 (DIN 12880)
- 4 potential-free relay outputs that can be activated via MCS controller
- Ethernet interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- 1 access port Ø 50 mm (1.97 inch), left side
- 1 stainless steel rack included
- 4 castors (with 2 brakes)
- BINDER test certificate
General Specifications
There are no General Specifications available.
Series Modals
BINDER - MKF 115
[+] View Model SpecificationsDepth | 850 mm |
Height | 1718 mm |
Width | 995 mm |
Weight | 280 kg |
Additional Model Specifications | Interior dimensions Width (mm): 600 Height (mm): 480 Depth (mm): 400 Interior volume (l): 115 Racks (number standard/max.): 1 / 4 Load per shelf (kg): 30 Permitted total load (kg): 60 Temperature data Temperature range (°C): -40 - 180 Temperature fluctuation (±K): 0,1 - 1,0 Temperature variation (±K): 0,1 - 1,3 Mean warm-up rate acc. to IEC 60068-3-5 (K/min.): 5,5 Mean cooling rate acc. to IEC 60068-3-5 (K/min.): 4,5 Warm-up time from -40 °C to 180 °C (min.): 45 Cooling down time from 180 °C to -40 °C (min.): 120 Max. heat compensation up to 25 °C (kW): 2,5 Climate data Temperature range (°C): 10 - 95 Temperature fluctuation (±K): 0,1 - 1,3 Humidity range (% RH): 10 - 98 Humidity fluctuation (± RH %): = 2,5 Dew point temperature range (°C): 5 - 94 Max. heat compensation to 25 °C / 90% RH (kW): 0,4 Electrical data IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 Hz (V): 400 V 3N~ Nominal power (kW): 4,2 Energy consumption 25 °C / 60% RH (kW) 1): 1,25 Noise level approx. (dB (A)): 62 |
BINDER - MKF 240
[+] View Model SpecificationsDepth | 946 mm |
Height | 1713 mm |
Width | 1130 mm |
Weight | 360 kg |
Additional Model Specifications | Interior dimensions Width (mm): 735 Height (mm): 700 Depth (mm): 443 Interior volume (l): 228 Racks (number standard/max.): 1 / 6 Load per shelf (kg): 30 Permitted total load (kg): 70 Temperature data Temperature range (°C): -40 - 180 Temperature fluctuation (±K): 0,1 - 1,5 Temperature variation (±K): 0,5 - 2,0 Mean warm-up rate acc. to IEC 60068-3-5 (K/min.): 5,0 Mean cooling rate acc. to IEC 60068-3-5 (K/min.): 5,0 Warm-up time from -40 °C to 180 °C (min.): 45 Cooling down time from 180 °C to -40 °C (min.): 98 Heat compensation, max. (W): 2800 Climate data Temperature range (°C): 10 - 95 Humidity range (% RH): 10 - 98 Humidity fluctuation (± RH %): 0,5 - 3,0 Dew point temperature range (°C): 5 - 94 Heat compensation, max. (W) 1): 300 Electrical data IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 Hz (V): 400 V 3N~ Nominal power (kW): 5,1 |
BINDER - MKF 720
[+] View Model SpecificationsDepth | 1173 mm |
Height | 2005 mm |
Width | 1613 mm |
Weight | 590 kg |
Additional Model Specifications | Interior dimensions Width (mm): 1200 Height (mm): 1020 Depth (mm): 600 Interior volume (l): 734 Shelves (number standard/max.): 1 / 11 Load per shelf (kg): 40 Permitted total load (kg): 160 Temperature data Temperature range (°C) 1): -40 - 180 Temperature fluctuation (± K): 0,1 - 0,5 Temperature variation (± K): 0,1 - 1,8 Mean warm-up rate acc. to IEC 60068-3-5 (°C/min.): 4,8 Mean cooling rate acc. to IEC 60068-3-5 (°C/min.): 4,8 Warm-up time from -40 °C to 180 °C (min.): 74 Cooling down time from 180 °C to -40 °C (min.): 118 Heat compensation, max. (W): 6500 Climate data Temperature range (°C): 10 - 95 Humidity range (% RH): 10 - 98 Humidity fluctuation (± RH %): = 2,5 Dew point temperature range (°C): 5 - 94 Heat compensation, max. (W): 1000 Electrical data IP protection class acc. to EN 60529: IP 20 Voltage (± 10%) 50 / 60 Hz (V): 400 3N~ Nominal power (kW): 11,0 Noise level (approx. dB(A)): 65 |