Advanced® Model 3300 Micro-Osmometer for Osmolality Measurement
Automated, single sample, designed to process a 20µL sample, and one-minute test time, allows the Model 3300 to provide the best accuracy and precision in the market today.
Note: The Advanced Instruments Model 3300 is no longer available for purchase. The Advanced Instruments Model 3320 Osmometer has replaced this product. The information presented is for existing users.
Active Questions & AnswersAsk a Question
Operator & Service manual?
Osmometer Variability
Osmometers Service ProvidersView All (5)
Documents & ManualsView All Documents
Features of Advanced® Model 3300
- 20 µL is ideal for applications where the sample quantity is limited.
- Internal diagnostics
- On-board statistical analysis
- Multi-language operation
- Automated calibration
- Hot-Line® Technical Service available 24/7
General Specifications
Depth | 13 in |
Height | 15.5 in |
Width | 10.5 in |
Power Requirements | 100 to 130 VAC (50-60 Hz) or 200 to 250 VAC (50/60 Hz). 60 W consumption |
Weight | 16 lbs |
Additional Specifications
Memory Backup: Integral lithium cell; 10-year life (typical); (not user-replaceable)
Sample Volume: 20 µL
Sample Capacity: Single sample
Readout: 2 line by 24-character vacuum fluorescent display
Units: mOsm/kg H2O
Range: 0 to 2000 mOsm/kg H2O
Resolution: 1 mOsm/kg H2O
Communications: DTE EIA-232/V.24 (RS-232); serial port, parallel printer port and optional barcode scanner
Performance at Reference Conditions
Linearity: Less than ±1% from a straight line between 0 and 850 mOsm/kg H2O; less than ±1.5% above 850 mOsm/kg H2O
Repeatability: 0 to 400 mOsm: ±2 mOsm/kg H2O (1 S.D.) 400 to 2000 mOsm: ±0.5% (1 S.D.)
Drift: Less than 1 mOsm/kg H2O per month
Performance Over Operating Conditions
Temperature Effects: Less than 1 mOsm/kg H2O for every 5°C (9°F) ambient temperature change
Operating Conditions
Temperature: 18°C to 35°C (64°F to 95°F)
Humidity: 5 to 80% relative humidity (non-condensing)
Storage Temp.: -40°C to +45°C (-40°F to +113°)
Start-up Time: 30 seconds from power-on
Test Time: 60 seconds
Shipping Weight: 11.4 kg (25.0 lbs.)
Warranty: One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers.
Certification: This product is a certified IVD, CE and ETL-USA instrument
Installation Class: I
Over-Voltage:
Category: II
Pollution Degree: 2
Moisture Protection: IPX0 (ordinary)