Advanced Instruments - Advanced® Model 3300

Manufactured by  Advanced Instruments
Follow this Equipment

Advanced® Model 3300 Micro-Osmometer for Osmolality Measurement

The Advanced Instruments Model 3300 Micro-Osmometer determines the osmolality of solutions using freezing point depression (FPD). It is ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications.  
  
Automated, single sample, designed to process a 20µL sample, and one-minute test time, allows the Model 3300 to provide the best accuracy and precision in the market today.  
  
Note: The Advanced Instruments Model 3300 is no longer available for purchase. The Advanced Instruments Model 3320 Osmometer has replaced this product. The information presented is for existing users.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked byMary Keo

avatar placemark

Asked byjklein02

Osmometers Service ProvidersView All (5)

Request Support
Features of Advanced® Model 3300
  • 20 µL is ideal for applications where the sample quantity is limited.  
  • Internal diagnostics  
  • On-board statistical analysis  
  • Multi-language operation  
  • Automated calibration  
  • Hot-Line® Technical Service available 24/7
General Specifications
Depth 13 in
Height15.5 in
Width 10.5 in
Power Requirements 100 to 130 VAC (50-60 Hz) or 200 to 250 VAC (50/60 Hz). 60 W consumption
Weight16 lbs
Additional Specifications

Memory Backup: Integral lithium cell; 10-year life (typical); (not user-replaceable)  
Sample Volume: 20 µL  
Sample Capacity: Single sample  
Readout: 2 line by 24-character vacuum fluorescent display  
Units: mOsm/kg H2O  
Range: 0 to 2000 mOsm/kg H2O  
Resolution: 1 mOsm/kg H2O  
Communications: DTE EIA-232/V.24 (RS-232); serial port, parallel printer port and optional barcode scanner  
  
Performance at Reference Conditions  
Linearity: Less than ±1% from a straight line between 0 and 850 mOsm/kg H2O; less than ±1.5% above 850 mOsm/kg H2O  
Repeatability: 0 to 400 mOsm: ±2 mOsm/kg H2O (1 S.D.) 400 to 2000 mOsm: ±0.5% (1 S.D.)  
Drift: Less than 1 mOsm/kg H2O per month  
  
Performance Over Operating Conditions  
Temperature Effects: Less than 1 mOsm/kg H2O for every 5°C (9°F) ambient temperature change  
  
Operating Conditions  
Temperature: 18°C to 35°C (64°F to 95°F)  
Humidity: 5 to 80% relative humidity (non-condensing)  
Storage Temp.: -40°C to +45°C (-40°F to +113°)  
Start-up Time: 30 seconds from power-on  
Test Time: 60 seconds  
Shipping Weight: 11.4 kg (25.0 lbs.)  
Warranty: One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers.  
Certification: This product is a certified IVD, CE and ETL-USA instrument  
Installation Class: I  
Over-Voltage:  
Category: II  
Pollution Degree: 2  
Moisture Protection: IPX0 (ordinary)