Thermo Scientific - Safe 2020
Manufactured by Thermo Scientific
Thermo Scientific Safe 2020 Class II cabinets are designed to deliver maximum safety, comfort and convenience.
Safe 2020 biological safety cabinets can be used wherever protection is of paramount importance in safety levels I, II and III, including research, pharmaceutical quality control labs, pharmacies, and clinics/hospitals. Combined with the Night-Set-Back mode (reduced blower speed when the window is closed), the Safe 2020 consumes hundreds of kilowatts less energy than competing cabinets. Not only is our Safe 2020 the preferred cabinet for enhanced containment, comfort, and convenience,?t it saves you money and reduces your lab's overall carbon footprint.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Biological Safety Cabinets Service ProvidersView All (5)
Documents & ManualsView All Documents
Features of Safe 2020
Featuring an advanced design, these cabinets comply with stringent regulatory standards and are certified to EN 12469. Not available in the U.S.
- Superior personnel and product protection with best-in-class air flow and filter technology
- Exceptionally comfortable ergonomic design prevents user fatigue and promotes safe working habits
- Maximized energy efficiency for environmental protection and significant cost savings
- Superior DC technology uses energy efficiently, emits less heat
- Night-Set-Back mode (reduced blower speed when the window is closed) saves energy
- Smart Flow* digital technology maintains consistent airflow during normal filter loading or temporary airflow obstruction
- Supply/exhaust air filter: H14 HEPA EN 1822, 99.995% at the most penetrating particle size (MPPS)
- Sash opening: 20cm working height; can be fully opened to 77.3cm for thorough cleaning
- Optional crossbeam UV irradiation in both side walls illuminates entire working area—ensures thorough disinfection for maximum sample protection
- Quiet: noise levels 56 or 58dbA, depending on model
Safety and Protection
- Pressure sensor monitoring ensures safe airflow across the entire work surface
- Independent alarm system immediately alerts to any airflow obstruction
Innovative Construction
- Frameless front sash and side glass window yields unobstructed view of the total work area
- Aerosol-tight window sealing for additional protection when cabinet is in stand-by mode
- Scratch-resistant stainless-steel work surface with indentations that trap spilled liquids
Ordering Alerts: Cabinets are fully configurable, with options and accessories to meet your specific needs. Contact your Thermo Scientific representative for more information.
Certifications: EN 12469
General Specifications
There are no General Specifications available.
Series Modals
Thermo Scientific - Safe 2020 0.9
[+] View Model SpecificationsAirflow (CFM) | 170 CFM |
Depth | 31.5 in |
Height | 62.2 in |
Internal Depth | 18.3 in |
Internal Height | 30.7 in |
Width | 39.4 in |
Internal Width | 35.4 in |
Power Requirements | 230V 50/60Hz, 160 W consumption |
Lighting | >800lx (74fc) |
Noise Level | 56 dB |
Weight | 375 lbs |
Shipping Weight | 419 lbs |
Shipping Dimensions | 69.3 |
Cabinet Type | Class II Type A2 |
Filter Type | HEPA |
Additional Model Specifications | Working height of front window: 8" Maximum opening height of front window: 30.4" Maximum load capacity of one-piece work surface: 110 lbs Filter: H14 HEPA EN 1822, 99.995% at the most penetrating particle size (MPPS) Load Capacity: 110 lb. (51kg) Outlets: One duplex on each side wall Port: 4 access (2 on each side wall) Sash Opening: 8 in. (20 cm); 30.4 in. (77.3cm) max. Service Fixtures: Up to 4 through access ports; up to 6 on rear wall Heat output: 160 W (546 BTU/hr) |
Thermo Scientific - Safe 2020 1.2
[+] View Model SpecificationsAirflow (CFM) | 227 CFM |
Depth | 31.5 in |
Height | 62.2 in |
Internal Depth | 18.3 in |
Internal Height | 30.7 in |
Width | 39.4 in |
Internal Width | 47.2 in |
Power Requirements | 230V 50/60Hz, 200 W consumption |
Lighting | >850lx (79fc) |
Noise Level | 56 dB |
Weight | 441 lbs |
Shipping Weight | 496 lbs |
Shipping Dimensions | 69.3 |
Cabinet Type | Class II Type A2 |
Filter Type | HEPA |
Additional Model Specifications | Working height of front window: 8" Maximum opening height of front window: 30.4" Maximum load capacity of one-piece work surface: 110 lbs Filter: H14 HEPA EN 1822, 99.995% at the most penetrating particle size (MPPS) Load Capacity: 110 lb. (51kg) Outlets: One duplex on each side wall Port: 4 access (2 on each side wall) Sash Opening: 8 in. (20 cm); 30.4 in. (77.3cm) max. Service Fixtures: Up to 4 through access ports; up to 6 on rear wall Heat output: 200 W (682 BTU/hr) |
Thermo Scientific - Safe 2020 1.5
[+] View Model SpecificationsAirflow (CFM) | 284 CFM |
Depth | 31.5 in |
Height | 62.2 in |
Internal Depth | 18.3 in |
Internal Height | 30.7 in |
Width | 63 in |
Internal Width | 59.1 in |
Power Requirements | 230V 50/60Hz, 265 W consumption |
Lighting | >1250lx (116fc) |
Noise Level | 58 dB |
Weight | 507 lbs |
Shipping Weight | 573 lbs |
Shipping Dimensions | 69.3 |
Cabinet Type | Class II Type A2 |
Filter Type | HEPA |
Additional Model Specifications | Working height of front window: 8" Maximum opening height of front window: 30.4" Maximum load capacity of one-piece work surface: 110 lbs Filter: H14 HEPA EN 1822, 99.995% at the most penetrating particle size (MPPS) Load Capacity: 110 lb. (51kg) Outlets: One duplex on each side wall Port: 4 access (2 on each side wall) Sash Opening: 8 in. (20 cm); 30.4 in. (77.3cm) max. Service Fixtures: Up to 4 through access ports; up to 6 on rear wall Heat output: 265 W (904 BTU/hr) |
Thermo Scientific - Safe 2020 1.8
[+] View Model SpecificationsAirflow (CFM) | 341 CFM |
Depth | 31.5 in |
Height | 62.2 in |
Internal Depth | 18.3 in |
Internal Height | 30.7 in |
Width | 74.8 in |
Internal Width | 70.9 in |
Power Requirements | 230V 50/60Hz, 340 W consumption |
Lighting | >1300lx (121fc) |
Noise Level | 58 dB |
Weight | 617 lbs |
Shipping Weight | 694 lbs |
Shipping Dimensions | 69.3 |
Cabinet Type | Class II Type A2 |
Filter Type | HEPA |
Additional Model Specifications | Working height of front window: 8" Maximum opening height of front window: 30.4" Maximum load capacity of one-piece work surface: 110 lbs Filter: H14 HEPA EN 1822, 99.995% at the most penetrating particle size (MPPS) Load Capacity: 110 lb. (51kg) Outlets: One duplex on each side wall Port: 4 access (2 on each side wall) Sash Opening: 8 in. (20 cm); 30.4 in. (77.3cm) max. Service Fixtures: Up to 4 through access ports; up to 6 on rear wall Heat output: 340 W (1161 BTU/hr) |