Forma* 1400 Series Class II, Type A2 Biological Safety Cabinets
Thermo Scientific cabinets feature an innovative design for a brighter, quieter, more comfortable environment. With its angled front, transparent glass sidewalls, and low noise level (<61dBA), the Thermo Scientific Forma 1400 Series ensures maximum operator comfort. Meets NSF Standard 49.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked byCC79
Biological Safety Cabinets Service ProvidersView All (5)
Documents & ManualsView All Documents
Features of Forma 1400 Series
- Microprocessor-based controller monitors and maintains airflow; displays read-out in real-time bar graph
- Intuitive user interface for easy programming of all key functions, including UV timer, auto-start and alarms
- Audible and visual alarms alert to unsafe condition
- Ergonomic angled front
- Transparent glass side walls and coated back wall
- Reduced noise levels
- Adjustable cabinet height
- Hinged-cabinet front and sliding sash
- Optional UV light is programmable; operation is disabled unless slash is fully closed
- Pushbutton motorized sash automatically stops at 8 or 10 in. (20.3 or 25.4cm) height, depending on model
Ordering Alerts: Optional accessories include UV light, manually adjustable support stand, electric support stand, armrest, exhaust transition, and foot switch.
Compliance: NSF/ANSI* 49
General Specifications
There are no General Specifications available.
Series Modals
Thermo Scientific - 4-foot Model
[+] View Model SpecificationsAirflow (CFM) | 361 CFM |
Depth | 35.3 in |
Height | 86 in |
Internal Depth | 22.9 in |
Internal Height | 27.6 in |
Width | 52.8 in |
Internal Width | 47.3 in |
Power Requirements | 115 volts, 60 Hz, 482 W consumption, 20 A supply circuitry, NEMA 5-20P plug, 2 GFCI protected duplex 120V (7 A total) outlets |
Lighting | Light intensity, foot candle (lux): >111 (>1200) |
Noise Level | 63 dB |
Weight | 496 lbs |
Cabinet Type | Class II Type A2 |
Additional Model Specifications | Spillage containment: 15 gal. (57 L) Air velocity, inflow: 100-110 FPM Air velocity, downflow: 50-60 FPM |
Thermo Scientific - 6-Foot Model
[+] View Model SpecificationsAirflow (CFM) | 543 CFM |
Depth | 35.3 in |
Height | 86 in |
Internal Depth | 22.9 in |
Internal Height | 27.6 in |
Width | 76.4 in |
Internal Width | 70.9 in |
Power Requirements | 115 V, 60 Hz, 745 W consumption, 20 A supply circuitry, NEMA 5-20P plug, 2 GFCI protected duplex 120V (7 total) outlets |
Lighting | Light intensity, foot candle (lux): >111 (>1200) |
Noise Level | 65 dB |
Weight | 661 lbs |
Cabinet Type | Class II Type A2 |
Additional Model Specifications | Spillage containment: 15 gal. (57 L) Air velocity, inflow: 100-110 FPM Air velocity, downflow: 50-60 FPM |