PHCBi - MCO-18AIC(UV), MCO-36AIC(UV)

Manufactured by  PHCBi
Follow this Equipment

Active Questions & AnswersAsk a Question

There are no current Discussions

CO2 Incubator Service ProvidersView All (28)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features

There are no Features available.

General Specifications
CO2 Incubator TypeAir Jacketed
Depth 28 in
Humidity95+/-5% (ambient temp. 37°C, CO2 5%)
Temperature Uniformity0.25 °C
Height35.4 in
Internal Depth 20.6 in
Internal Height 26.2 in
Width 24.4 in
Internal Width 19.3 in
Power Requirements 115V, AC, 60Hz, NEMA 5-15
Chamber Volume6 cu. ft.
CO2 SensorInfrared (IR)
CO2 Range0 to 20 %
CO2 Accuracy0.1 %
Shelf Dimensions17.7
ConstructionInterior surface: inCu saFe™ copper-enriched stainless steel alloy for germicidal protection. Exterior cabinet: polyester finished, baked-on zinc galvanized steel, reinforced for stacking. Inner door: tempered glass with positive latch. Outer door: left hand swing standard, reversible to right hand swing
Incubator Orientation Benchtop
Additional Specifications

Heating System
Method: Patented Direct Heat and Air Jacket™
Heating Elements: 314 watts per chamber
Temperature Control: microprocessor controlled P.I.D.
Digital Temperature Display: resolution 0.1ºC

CO2 System
CO2 System Electronics: microprocessor P.I.D./R
CO2 Setpoint and Resolution: ±0.1%
CO2 Inlet Connect/Pressure: 5 PSIG, 0.03MpaG
CO2 Switchover System: optional

Humidification System
Method: natural evaporation
Water Level Sensor: optical, low water level alarm
Exterior dimensions, stacked: 30.3"W x 27.9"F-B x 70.8"H

Contamination Control
SafeCell™ UV System: standard
inCu saFe™ Interior: standard
Decontamination Method: programmable UV sterilization of air and water pan
UV Lamp: 4 watt, 253.7 nanometer narrow bandwidth, ozone-free
Microbiological Filters: 0.3 microns, 99.97% efficient, on air and CO2 inputs

Control, Alarm, Monitoring and Electrical
Microprocessor Control: SANYO-built electronic components
Control Position: door-mounted, eye-level location on dual stacked configuration
Alarm System: Overtemp., CO2 and temperature deviation, low water level, door ajar, lamp failure alarm
Remote Alarm Contacts: 30V DC, 2 amps allowable
Communications (Optional): MCO-420MA data port available, 4 to 20 MA signal