BWB Technologies USA LLC - BIO Flame Photometer For Biological Application

Manufactured by  BWB Technologies USA LLC
Follow this Equipment

Purpose Built Flame Photometer with specialized software, firmware and hardware.

Purpose Built Flame Photometer with specialized software, firmware and hardware.

Product features:

  • Specially configured for a range of user interests
  • IRS(International Reference Standard) caesium or lithium option
  • Single point calibration correction  
    International reference standards
  • 5 and 6 point multi-ion calibration standards for two separate calibration curves
  • Analysing Li, and CA along with NA and K
  • Operator independent determination of results
  • Advanced automated determination of results
  • Integrated printer with environmentally friendly ink
  • IQ, OQ, PQ web-based certification standard on BIO 943 and available on all BWB fes units
  • Correction of CA for the interference from high levels of NA 

The BWB BIO has been created for measuring biological  solutions and is modernized for today’s research work;  which requires routine, quick measurements, with high levels of accuracy for lithium, calcium, sodium and potassium elements.

Our recent release is the combination of several “wish lists” from PhD’s in Germany, Chile, Colombia, USA and Canada principally in renal research, diabetes and Alzheimer’s research. Production reagents and solutions QA and QC are facilitated quickly by our unique simultaneous detection and data processing for all atoms of interest.

When connected to a PC with the BWB FP-PC app, the user can create .pdf files, .csv files, data presentations and reports that include all the key analysis data. The FP-PC Software also allows setup of the Automatic Fluid Handling System, AFHS, originally designed to work specifically with the BIO unit.

The software enables you to view and edit the calibration curves. The intuitive user-friendly operation of the BWB BIO spans all languages and appeals to all levels of users. The BWB BIO is the must-have instrument in today’s research laboratory.

For sites that do not allow PCs or data acquisition in the test lab, the Printer provides the “hard copy” required for QA and QC purposes. The environmentally friendly print ink used by the printer has a minimum character capture life of 10 years. The printer is also compatible with other readily available paper stock.

To enhance accuracy and reproducibility there is an option to use either Caesium or Lithium as an Internal Reference System (IRS) that acts as an internal standard to adjust small changes during calibration and measuring.

Active Questions & AnswersAsk a Question

There are no current Discussions

Elemental Analyzers Service ProvidersView All (6)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features

There are no Features available.

General Specifications

There are no General Specifications available.

Additional Specifications

Sample rate: 2-3.5ml/min   
   
Tubing materials: Silicone and Tygon®    
   
Specificity: Na/K/Li = <0.5% to each other when   
   
equal in concentration at <100ppm   
   
   
Required desk space for Instrument :   
   
50cm (H) x 45cm (W) x 45cm (D)   
   
For Safety reasons the Flame photometer requires1m of unobstructed space above to allow dissipation of heat from the chimney.   
   
   
Instrument size:   
   
BIO Flame Photometer   
51cm (H) x 38cm (W) x 41cm (D)   
(20in x 14in x 16in)   
   
   
Shipping:   
62cm (H) x 55cm (W) x 47cm (D)   
(24in x 22in x 19in)   
   
   
Weight:   
   
BIO Flame Photometer 15.3kg   
   
Shipping  BIO Flame Photometer 25kg   
   
   
   
Optimal range:    
   
Multi-point calibration   
   
Na  0.0043  300mmol/l*   
   
K  0.0013  200mmol/*    
   
Li  0.0072  20mmol/l*   
   
Ca  0.025  20mmol/l*   
   
*With required Dilution Ratios   
   
   
Reproducibility:   
   
<1% Coefficient of variability for 20 consecutive samples over ten minutes   
at concentrations of 100ppm equivalent   
or less. (After instrument stabilisation).   
   
   
Limit of detection (LOD) and   
limit of quantification (LOQ):   
   
LOD LOQ    
   
Na  0.0013mmol/l Na  0.0043mmol/l    
   
K  0.0005mmol/l K  0.0013mmol/l    
   
Li  0.003mmol/l Li  0.0072mmol/l    
   
Ca  0.0075mmol/l Ca  0.025mmol/l    
   
   
Time to stability:   
   
Less than 15 seconds after sample   
is introduced into the flame   
   
   
Ions measured:    
   
Serum Na and K or Li and Ca   
Urine Na and K or Li and Ca   
   
Calibration concentrations:    
   
   
URINE    
   
Point 1 Point 2 Point 3   
Na 25mmol/l Na 50mmol/l Na 100mmol/l   
K 25mmol/l K 50mmol/l K 100mmol/l   
Li 2.5mmol/l Li 5mmol/l Li 10mmol/l   
Ca 2.5mmol/l Ca 5mmol/l Ca 10mmol/l   
   
   
Point 4 Point 5    
Na 200mmol/l Na 300mmol/l    
K 150mmol/l K 200mmol/l    
Li 15mmol/l Li 20mmol/l    
Ca 15mmol/l Ca 20mmol/l    
   
   
SERUM    
   
Point 1 Point 2 Point 3   
Na 120mmol/l Na 130mmol/l Na 140mmol/l   
K 3mmol/l K 4mmol/l K 5mmol/l   
Li 0.25mmol/l Li 0.5mmol/l Li 1mmol/l   
Ca 1.5mmol/l Ca 2mmol/l Ca 2.5mmol/l   
   
   
Point 4 Point 5    
Na 150mmol/l Na 160mmol/l    
K 6mmol/l K 7mmol/l    
Li 1.5mmol/l Li 2mmol/l    
Ca 3mmol/l Ca 3.5mmol/l   
   
   
Recommended minimum warm up   
time based on ambient temperature:   
   
21°C  40 minutes    
   
   
Power requirements: 100V  250V AC at 50 or 60Hz   
   
Fuel requirement:   
   
Propane, Butane or Natural Gas*   
regulated to 19Bar. Flow rate of 0.4l/min   
   
*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.    
   
   
Readout:    
   
LCD, four line, alpha numeric, back lit.