Product features:
- Specially configured for a range of user interests
- IRS(International Reference Standard) caesium or lithium option
- Single point calibration correction
International reference standards - 5 and 6 point multi-ion calibration standards for two separate calibration curves
- Analysing Li, and CA along with NA and K
- Operator independent determination of results
- Advanced automated determination of results
- Integrated printer with environmentally friendly ink
- IQ, OQ, PQ web-based certification standard on BIO 943 and available on all BWB fes units
- Correction of CA for the interference from high levels of NA
The BWB BIO has been created for measuring biological solutions and is modernized for today’s research work; which requires routine, quick measurements, with high levels of accuracy for lithium, calcium, sodium and potassium elements.
Our recent release is the combination of several “wish lists” from PhD’s in Germany, Chile, Colombia, USA and Canada principally in renal research, diabetes and Alzheimer’s research. Production reagents and solutions QA and QC are facilitated quickly by our unique simultaneous detection and data processing for all atoms of interest.
When connected to a PC with the BWB FP-PC app, the user can create .pdf files, .csv files, data presentations and reports that include all the key analysis data. The FP-PC Software also allows setup of the Automatic Fluid Handling System, AFHS, originally designed to work specifically with the BIO unit.
The software enables you to view and edit the calibration curves. The intuitive user-friendly operation of the BWB BIO spans all languages and appeals to all levels of users. The BWB BIO is the must-have instrument in today’s research laboratory.
For sites that do not allow PCs or data acquisition in the test lab, the Printer provides the “hard copy” required for QA and QC purposes. The environmentally friendly print ink used by the printer has a minimum character capture life of 10 years. The printer is also compatible with other readily available paper stock.
To enhance accuracy and reproducibility there is an option to use either Caesium or Lithium as an Internal Reference System (IRS) that acts as an internal standard to adjust small changes during calibration and measuring.
Active Questions & AnswersAsk a Question
There are no current Discussions
Elemental Analyzers Service ProvidersView All (6)
Documents & Manuals
There are no Documents or Manuals available.
Features
There are no Features available.
General Specifications
There are no General Specifications available.
Additional Specifications
Sample rate: 2-3.5ml/min
Tubing materials: Silicone and Tygon®
Specificity: Na/K/Li = <0.5% to each other when
equal in concentration at <100ppm
Required desk space for Instrument :
50cm (H) x 45cm (W) x 45cm (D)
For Safety reasons the Flame photometer requires1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size:
BIO Flame Photometer
51cm (H) x 38cm (W) x 41cm (D)
(20in x 14in x 16in)
Shipping:
62cm (H) x 55cm (W) x 47cm (D)
(24in x 22in x 19in)
Weight:
BIO Flame Photometer 15.3kg
Shipping BIO Flame Photometer 25kg
Optimal range:
Multi-point calibration
Na 0.0043 300mmol/l*
K 0.0013 200mmol/*
Li 0.0072 20mmol/l*
Ca 0.025 20mmol/l*
*With required Dilution Ratios
Reproducibility:
<1% Coefficient of variability for 20 consecutive samples over ten minutes
at concentrations of 100ppm equivalent
or less. (After instrument stabilisation).
Limit of detection (LOD) and
limit of quantification (LOQ):
LOD LOQ
Na 0.0013mmol/l Na 0.0043mmol/l
K 0.0005mmol/l K 0.0013mmol/l
Li 0.003mmol/l Li 0.0072mmol/l
Ca 0.0075mmol/l Ca 0.025mmol/l
Time to stability:
Less than 15 seconds after sample
is introduced into the flame
Ions measured:
Serum Na and K or Li and Ca
Urine Na and K or Li and Ca
Calibration concentrations:
URINE
Point 1 Point 2 Point 3
Na 25mmol/l Na 50mmol/l Na 100mmol/l
K 25mmol/l K 50mmol/l K 100mmol/l
Li 2.5mmol/l Li 5mmol/l Li 10mmol/l
Ca 2.5mmol/l Ca 5mmol/l Ca 10mmol/l
Point 4 Point 5
Na 200mmol/l Na 300mmol/l
K 150mmol/l K 200mmol/l
Li 15mmol/l Li 20mmol/l
Ca 15mmol/l Ca 20mmol/l
SERUM
Point 1 Point 2 Point 3
Na 120mmol/l Na 130mmol/l Na 140mmol/l
K 3mmol/l K 4mmol/l K 5mmol/l
Li 0.25mmol/l Li 0.5mmol/l Li 1mmol/l
Ca 1.5mmol/l Ca 2mmol/l Ca 2.5mmol/l
Point 4 Point 5
Na 150mmol/l Na 160mmol/l
K 6mmol/l K 7mmol/l
Li 1.5mmol/l Li 2mmol/l
Ca 3mmol/l Ca 3.5mmol/l
Recommended minimum warm up
time based on ambient temperature:
21°C 40 minutes
Power requirements: 100V 250V AC at 50 or 60Hz
Fuel requirement:
Propane, Butane or Natural Gas*
regulated to 19Bar. Flow rate of 0.4l/min
*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.
Readout:
LCD, four line, alpha numeric, back lit.