BWB Technologies USA LLC - BWB Flash Flame Photometer
The “Flash” BWB Flame Photometer mini product line was created in January 2018 to answer the request from global...
The design criteria for this new 2018 line was to provide a modernized replacement flame photometer for the hundreds of units in service in the USA and Canada that are now decades old. These older units have no replacement parts and no servicing is currently available. These are Perkin Elmer, Jenway Instruments, Corning and Sherwood Instruments units as well as the units supplied by Cole Parmer, VWR and Fisher Scientific.
When the costs of adding the extra components that are required to operate the flame photometers sold by our competition are totalled up, (calibration solutions, diluent solutions, various user replacement optional detection filters, optional software that ties to computers; AND the big cost, the required Air Compressor) the benefit of simplicity and cost savings of the “Just Add Gas” - "only one order number needed" we provide is apparent.
Simple to order, simple to use, minimal tabletop space required, Major Power in simple software as standard (FP-PC), the new "FLASH" is another exciting offering from BWB Technologies USA. For new emerging markets: aqua farming, flow injection analyses, educational markets as well as the Replacement Flame Photometer marketplace in the USA.
Active Questions & AnswersAsk a Question
There are no current Discussions
Elemental Analyzers Service ProvidersView All (6)
Documents & Manuals
There are no Documents or Manuals available.
Features
There are no Features available.
General Specifications
Depth | 47 cm |
Height | 62 cm |
Width | 55 cm |
Weight | 15.3 kg |
Additional Specifications
Technical Data
Sample rate: 2-3.5ml/min
Tubing materials: Silicone and Tygon®
Required desk space for Instrument: 50cm (H) x 45cm (W) x 45cm (D)*
For safety reasons, a Flame photometer requires 1m
of unobstructed space above to allow dissipation of
heat from the chimney.
Instrument size: Flash Flame Photometer, 51cm (H) x 38cm (W) x 41cm (D), (20in x 14in x 16in)
Shipping: 62cm (H) x 55cm (W) x 47cm (D), (24in x 22in x 19in)
Weight: Flash Flame Photometer – 15.3kg (33.7lbs), Shipping – 22kg (48.5 lbs)
Optimal range:
Single point calibration
Na 0.1 – 60ppm
K 0.05 – 100ppm
Li 0.03 – 30ppm
Ca 1.0 – 100ppm
Multi Point Calibration
Na 0.1 – 1000ppm
K 0.05 – 1000ppm
Li 0.05 – 1000ppm
Ca 1.0 – 1000ppm
Reproducibility:
<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilization).
Chemical Interference:
Na/K/Li = <0.5% to each other when
equal in concentration at <100ppm
Limit of detection (LOD) and
limit of quantification (LOQ)
LOD LOQ
Na – 0.03 ppm Na – 0.1 ppm
K – 0.02 ppm K – 0.05 ppm
Li – 0.02 ppm Li – 0.05 ppm
Ca – 0.3 ppm Ca – 1.0 ppm
Time to stability:
Less than 15 seconds after the sample is introduced into the flame
Ions measured:
Potassium
Sodium
Lithium
Calcium
Interfaces:
USB
Optional Integrated Printer
.CSV and PDF generated repors and files via FP_PC software
Recommended minimum warm up
time-based on ambient temperature: 21ºC – 40 minutes
Power requirements: 100V – 250V AC at 50 or 60Hz automatically selecting
Fuel requirement:
Propane, Butane or Natural Gas*
regulated to 19Bar. Flow rate of 0.4l/min
*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.
Readout:
LCD, four line, alpha numeric, back lit.