Thermo Scientific - Fluoroskan Ascent FL

Manufactured by  Thermo Scientific
Follow this Equipment

Combined microplate fluorometer and luminometer for a variety of applications

The Thermo Scientific Fluoroskan® Ascent FL is a computer-controlled microplate fluorometer and luminometer for both luminometric and fluorometric research applications such as protein and enzyme studies, molecular interactions, nucleic acid quantification, reporter gene, fluorometric kinase, immuno and cell based assays.  
  
Fluoroskan Ascent FL is a compact and robust instrument with excellent optical performance for a variety of fluorometric and luminometric research applications offering versatile plate formats, fast reading speeds, up to three dispensers and top / bottom reading of plates.  
  
5210450: Fluoroskan Ascent FL 100-240V, 50/60Hz   
5210460: Fluoroskan Ascent FL 100 - 240V, 50/60Hz with one dispenser   
5210462: Fluoroskan Ascent FL 100-240V, 50/60Hz with two dispensers   
5210463: Fluoroskan Ascent FL 100-240V, 50/60Hz with three dispensers
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked byilindb

avatar placemark

Asked byJpark

Microplate Readers Service ProvidersView All (12)

Request Support
Features of Fluoroskan Ascent FL

Fast reading speed  

  • Reads a 96-well plate in just 15 seconds, which is essential for kinetic applications  
  • Great for a wide variety of applications, such as cytotoxicity, ion channel, FRET and BRET  
  • Combination of fluorometric GFP measurement and a lumometric luciferase measurement from the same well is fast due to the versatility of assay programming  

Onboard dispensers for exact follow-up of kinetic reactions  

  • Equipped with up to three reagent dispensers for rapid kinetic assays, such as Ca2+ flux  
  • Supports simultaneous dispensing and reading, enabling monitoring of kinetic measurements from the very start of the reaction  
  • Very low dead volume and the backflush capability allow conservation of expensive reagent  

High sensitivity for both top and bottom reading  

  • Fiberless direct illumination optics for both top and bottom reading  
  • High sensitivity, wide dynamic range and low cross talk ensure accurate and precise results for 1- to 384-well plates  

Specially designed for automation  

  • Reads up to 384-well plates for increased assay throughput  
  • Easily integrated with automated systems  

Includes: PC Software and filter pairs: Excitation: 355nm/Emission: 460nm, Excitation: 485nm/ Emission: 538nm   Warranty and Service Offering: 1 year   Recommended for: Ca2+ flux assays; Cell proliferation; Cytotoxicity; Cell adhesion; DNA quantitation; Reporter gene assays; Hybridization assays; Quantitation of PCR products; FRET assays; BRET assays; Molecular beacon assays; Immunoassays; Enzyme activity; Bacterial quantitation; Phagocytosis; Oligonucleotide assays

General Specifications

There are no General Specifications available.

Additional Specifications

Note: 3 optional dispensers add 7.7 lbs to the weight.  
  
Fluorometry  
Excitation wavelength range: 320-700 nm  
Emission wavelength range: 360 - 670 nm  
Excitation filters: Up to eight in the excitation filter wheel. 355 nm and 485 nm filters included as standard. Other filters available upon request.  
Emission filters: Up to six filters in the emission filter wheel.  
460 nm and 538 nm filters included as standard. Other filters available upon request.  
Sensitivity: 2 fmol fluorescein/well in a black 96-well plate  
Dynamic range: >6 decades  
  
Luminometry  
Spectral range: 270 - 670 nm  
Sensitivity: 40 amol ATP/well using flash reaction, white 384-well plate  
Dynamic range: > 9 decades over whole gain setting area  
  
Dispensing  
Number of dispensers: Up to 3  
Dispensing volume: 1-1000 µl in 1 µl increments.  
Dispensing speed: 25 s, 96-well plate, 5 µl/well  
Accuracy: ±3 µl average for volumes 5 µl and above  
Precision: 5-19 µl < 5%, 20-1000 µl < 2%   
  
General features  
Sample format: 1 - 384-well plates  
Measurement speed: 15 s, 96-well plate  
Wavelength selection: Filters  
Light source: Quartz-halogen lamp  
Detector: Photomultiplier tube  
Incubator: From ambient +3°C to 45°C, at ambient 25°C  
Shaking: Orbital shaker  
User interface: Requires, but does not include a personal computer  