Multiskan GO has been designed to ensure first-rate performance and high quality results with minimal user effort. Extensive automatic self diagnostics at instrument start-up verifies all major instrument functionalities and continuous runtime control of optical and mechanical functions guarantee stable day-to-day and year-on-year performance and reliability.
- Freely selectable wavelengths from 200 – 1000 nm for the demands of various assays
- Both microplate and cuvette reading for any throughput requirements
- Very fast plate measurements and a full spectrum of a sample in less than 10 seconds
- High quality data guaranteed by extensive self diagnostics
- A unique power save function for reduced energy consumption
- Visual internal software on a large color screen for quick measurements
- Easy and logical assay setup for demanding assays with the powerful SkanIt Software
- A selection of eight different languages
Typical applications
- DNA and RNA quantitation and purity
- Protein assays
- Enzyme assays
- Kinetic assays
- Immunoassays
- Cell proliferation and cytotoxicity assays
- Apoptosis assays
- Reporter gene assays
- GPCR assays
General Specifications
There are no General Specifications available.
Additional Specifications
Wavelength selection: Monochromator
Light source: Xenon flash lamp
Wavelength range: 200 -1000 nm with 1 nm steps
Read-out range: Up to 4 Abs
Bandwidth: 2 nm
Plate
Linearity @ 450nm: 0 - 2.5 Abs, ± 2% (96-well plate)
Accuracy @ 450nm: 1.0% + 0.003 Abs (0 - 2.0 Abs) 2.0% (2.0 - 2.5 Abs)
Precision @ 450nm: CV < 1.0%
Plate types: 96- and 384-well plates
Measurement speed (from A1 back to A1): 6 s with 96-well plate 10 s with 384-well plate
Shaking: Linear
Cuvette
Linearity @ 450nm: 0 - 2.5 Abs, ± 2%
Accuracy @ 450nm: 1.0% + 0.003 Abs (0 - 2.0 Abs) 2.0% (2.0 - 2.5 Abs)
Precision @ 450nm: CV < 1.0%
Cuvette types: Standard, micro and ultramicro cuvettes, TrayCell
General Specifications
Spectral scanning speed: 10 s from 200 to 1000 nm with 1 nm steps
Incubation: From ambient + 4°C to 45°C
Internal user interface: 4.5´´ - color display USB memory stick position for data export USB port for external printer (HP PCL5 compatible)
PC control: SkanIt Software using USB connection
Max power consumption: 90 W
Power save consumption: < 5 W
Series Modals
Thermo Scientific - 51119200
[+] View Model SpecificationsBandwidth | 2 nm |
Depth | 16.9 in |
Height | 10.2 in |
Width | 11.2 in |
Power Requirements | 100 - 240 V (50/60 Hz), 90 W consumption, < 5 W save consumption |
Weight | 23.8 lbs |
Scanning Speed | 10 s from 200 to 1000 nm with 1 nm steps |
Light Source | Xenon flash lamp |
Additional Model Specifications | Wavelength selection: Monochromator Read-out range: Up to 4 Abs Linearity @ 450nm: 0 - 2.5 Abs, ± 2% (96-well plate) Accuracy @ 450nm: 1.0% + 0.003 Abs (0 - 2.0 Abs) 2.0% (2.0 - 2.5 Abs) Precision @ 450nm: CV < 1.0% Plate/ Cuvette types: 96- and 384-well plates Measurement speed: 6 s with 96-well plate, 10 s with 384-well plate Shaking: Linear Internal user interface: 4.5´´ - color display USB memory stick position for data export USB port for external printer (HP PCL5 compatible) PC control: SkanIt Software using USB connection |
Thermo Scientific - 51119300
[+] View Model SpecificationsBandwidth | 2 nm |
Depth | 16.9 in |
Height | 10.2 in |
Width | 11.2 in |
Power Requirements | 100 - 240 V (50/60 Hz), 90 W consumption, < 5 W save consumption |
Weight | 23.8 lbs |
Scanning Speed | 10 s from 200 to 1000 nm with 1 nm steps |
Light Source | Xenon flash lamp |
Additional Model Specifications | Wavelength selection: Monochromator Read-out range: Up to 4 Abs Linearity @ 450nm: 0 - 2.5 Abs, ± 2% Accuracy @ 450nm: 1.0% + 0.003 Abs (0 - 2.0 Abs) 2.0% (2.0 - 2.5 Abs) Precision @ 450nm: CV < 1.0% Plate/ Cuvette types: Standard, micro and ultramicro cuvettes, TrayCell Internal user interface: 4.5´´ - color display USB memory stick position for data export USB port for external printer (HP PCL5 compatible) PC control: SkanIt Software using USB connection |