VELP Scientifica - SER 148 Series
Solvent extraction system suitable for the separation of a substance or a group of elements from solid and semi-solid samples.
Up to 5 times faster than traditional Soxhlet (hot solvent vs. cold solvent)
Low solvent consumption (high solvent recovery, up to 75%)
Limited cost per analysis
Convenience & Choice
Select from 29 on-board empty programs
Two models available to match sample load
SER 148/3, a 3-position solvent extraction system
SER 148/6, a 6-position solvent extraction system
Operator Safety
IP 55 protection from particles and liquids
Solvent Extractor - Applications and HU6 Hydrolysis Unit
Solvent extraction is used for determination of different substances present in products related to agriculture, industry or in environmental samples.
SER Series can be used, according to official methods, for determining fats, tensides, plasticizers and pesticides on many matrices, including food and feed samples, detergents, rubber and plastic formulates, and many others.
No sample transfer, coming from HU6 Hydrolysis Unit, prevents any possible sample loss, as crucibles can also be used on the two units.
Solvent Extractors - Randall Method, Faster Results
Perform single or multiple solvent extractions with Randall method, an improvement of Soxhlet technique, as hot solvents are used to speed up the analysis.
In fact, the fast solubilization made by hot solvent permits a considerable reduction of the extraction time (approx. 90 minutes).
The extraction is made by immersion of the sample in the boiling solvent, followed by a rinsing phase with cold solvent.
Maximum reproducibility is ensured, as well as a high recovery rate of the solvent used.
SER Series can be used with a wide range of sample types and with a variety of solvents, for increased flexibility.
Active Questions & AnswersAsk a Question
There are no current Discussions
Liquid Handling Service ProvidersView All (9)
Documents & Manuals
There are no Documents or Manuals available.
Features
There are no Features available.
General Specifications
There are no General Specifications available.
Additional Specifications
GENERAL FEATURES
Construction material: epoxy painted stainless steel structure
Number of samples: SER 148/3 3 positions
SER 148/6 6 positions
2 Displays: working temperature / settable parameters
Dimensions (WxHxD): SER 148/3 480x620x390 mm (18.9x24.4x15.4 in)
SER 148/6 700x620x390 mm (27.6x24.4x15.4 in)
Weight: SER 148/3 30 Kg (66 lb)
SER 148/6 40 Kg (88 lb)
Power: SER 148/3 500 W
SER 148/6 950 W
PERFORMANCE
Reproducibility (RSD): ? 1%
Solvent recovery: from 50 to 75%
Max volume extraction cup: 150 ml
Number of programs: 29
Working temperature: from 100 to 260 °C
CONSUMPTION
Cooling water: 2 l/min
Sample quantity: from 0,5 to 15 g (generally 2-3 g)
Solvent volume: from 30 to 100 ml
SETTABLE PROGRAMS AND VALUES
Immersion time: from 0 to 999 minutes
Washing time: from 0 to 999 minutes
Recovery time: from 0 to 999 minutes