Anton Paar - TTK 600

Manufactured by  Anton Paar
Follow this Equipment

The TTK 600 Low-Temperature Chamber is a non-ambient attachment for powder X-ray diffraction studies from -190°C to 600°C.

The TTK 600 Low-Temperature Chamber is a non-ambient attachment for powder X-ray diffraction studies from -190°C to 600°C. Different sample holders allow investigating samples in reflection and transmission geometry. Samples can be measured in vacuum, air or inert gases. An antechamber option allows safe transfer of air-sensitive samples into the TTK 600. 
 
The TTK 600 fits to all common powder diffractometers and is the instrument of choice for X-ray structure analysis of various sample types at low and medium temperatures.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

X-ray Diffraction Service ProvidersView All (9)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of TTK 600

Fast and accurate temperature control 

  • Liquid nitrogen cooling to -190°C or compressed air cooling to -10°C * Resistance heating for rapid heating up to 600°C * Pt100 close to the sample for precise temperature measurement * Flexible liquid nitrogen hoses for movability even at -190°C * Improved heat transfer from heater to sample holder. Additional convection heating for the capillary sample holder and the transmission sample holder for reduced temperature deviation 

    Optimized for maximum sample flexibility 

    Various types of sample holders: * Standard sample holder for the investigation of powders in reflection geometry (different depths available) * Capillary sample holder for the investigation of powders in transmission geometry * Transmission sample holder for the investigation of foils, pastes and powders in transmission mode * Zero background holder for transparent samples * Integrated beam knife to minimize background at low 2Θ * Easy exchange and mounting of samples

General Specifications

There are no General Specifications available.

Additional Specifications

Operating temperature  
-190 °C to 600 °C with liquid nitrogen cooling 
-10 °C to 600 °C with compressed-air cooling 
 
Temperature measurement  
Pt 100 
 
Atmosphere  
Vacuum (10-2 mbar), air, inert gas 
 
Angle of incidence (2Θ)  
0 to 164 ° (standard sample holder) 
 
Max. operating pressure  
2 bar above atmospheric pressure 
 
Sample holder material  
Chromium-plated copper 
 
Sample size  
14 x 10 mm