Features of 260D Syringe Pump
260D pump has 1/8" Valco fittings for easy connection in process and pilot-plant plumbing. * Integral button-style pressure transducer in cylinder cap assures excellent stability and repeatability. * Rugged single-ratio drive train has auto-lubricating gears to assure long life and low maintenance. * Each ISCO D-Series syringe pump is bench tested at the factory. * Cylinder wash system protects the seals from deposits. * Exclusive "Smart Key" controller. All ISCO D-series pumps use the same controller that can operate any four pump modules, either independently or together. Operating modes include:
Single pump independent constant flow or pressure Single pump flow - or pressure - ramp programming Single pump dispensing Dual pump continuous constant flow or pressure Dual pump concentration gradients Four-pump configured systems * RS-232/485 serial interface for computer control or monitoring of operating parameters. * Redundant safety is designed in. All D-Series pumps have mechanical and electrical fail-safes, in addition to the factory default or user-set limits. * Corrosion resistance. Nitronic 50 - the standard material for cylinders, pistons, and caps - is an austenitic stainless steel with outstanding strength and corrosion resistance. * Standard seals are heavy-duty, graphite-impregnated Teflon, providing very long life under harsh operating conditions.
General Specifications
There are no General Specifications available.
Additional Specifications
Capacity: ?266 mL
?Flow Range: ?0.001 to 107 mL/min
?Flow Accuracy: ?0.5% of setpoint
?Displacement Resolution: ?16.6 nL/step
?Pressure Range: 0.7 to 517 bar (10 to 7,500 psi)
?Standard Pressure Accuracy: 0.5% Full Scale
?Optional Pressure Accuracy: ?0.1% Full Scale
?Wetted Materials: Nitronic 50, PTFE, Hastelloy C-276
?Plumbing Ports: ?1/8" Valco
?Operating Temperature: ?5 to 40 °C Ambient
?Power Requirements: 100 VAC, 117 VAC, 234 VAC, 50/60 Hz
?Dimensions (H x W x D): ?101 x 27 x 47 cm (39.8 x 10.7 x 18.4 in)
?Weight: ?Pump - 32.8 kg (72.3 lb)
Controller - 3.0 kg (6.5 lb)