HORIBA - Auto Titration Injector F-3005 and F-3006

Manufactured by  HORIBA
Follow this Equipment

The AutoTitration Injector is used for running automated and batch titrations with precisely controllable and programmable aliquot sizes and mixing...

The AutoTitration Injector is used for running automated and batch titrations with precisely controllable and programmable aliquot sizes and mixing speeds. 
 
The left syringe is used to inject aliquots of liquid into the sample to be scanned. The right syringe is used to mix the sample to be scanned, or to extract aliquots from the sample. 
 
Data are shown as a three-dimensional plot, using the z-axis to represent total volume of titrant delivered 
 
System Operation 
The F-3005 and F-3006 AutoTitrator are fully operated through FluorEssence™ software. 
 
There are two ways to alter the operation of the injector and syringes: 
 
To adjust the basic injector parameters: 
 
  • Volume of the syringes 
  • How fast to aspirate the syringes 
  • Filling and emptying the left syringe 
     
  • To modify a batch titration experiment: 
     
  • How many aliquots to add 
  • What volumes they ought to be 
  • Whether to mix the sample via the right syringe
  • Active Questions & AnswersAsk a Question

    There are no current Discussions

    Need Equipment Support?

    Request Support
    Documents & Manuals

    There are no Documents or Manuals available.

    Features of Auto Titration Injector F-3005 and F-3006
    • Automated and batch titrations * Controllable and programmable aliquot sizes and mixing speeds * Sample preparation and experiments time-saver * Eliminate technician to technician method variability
    General Specifications

    There are no General Specifications available.

    Additional Specifications

    Accuracy: 
    within ±1% 
     
    Precision: 
    within +0.2% 
     
    Resolution: 
    0.1% of syringe volume 
     
    Vol. increment: 
    0.1 to 100% of total syringe vol. 
     
    Speed: 
    1 to 250 seconds per full syringe stroke 
     
    Syringe Drive Mechanism: 
    stepper motor driven high precision lead screw with encoder 
     
    Electrical Specifications: 
    F-3005: 110 VAC at 1.4 A and 60 Hz; F-3006: 220 VAC at 0.7 A and 50 Hz 
     
    Dimensions (cm): 
    33.3 H x 15.6 W x 17.8 D 
     
    Communication interface: 
    RS-232 
     
    Computer requirements: 
    973005 from HORIBA Jobin Yvon or equivalent computer with free RS-232 port 
     
    Software requirements: 
    . Windows® 2000 or XP Pro  
    . FluorEssence™, version 2.1 or higher 
     
    Instrument compatibility with: 
    . FluoroLog®-3, FluoroMax®-4, and FluoroMax®-3 
     
    Certification: 
    CE

    Looking for New or Used Equipment?

    Shop on LabX

    Shop on Labx.com