HORIBA - Aqualog®

Manufactured by  HORIBA
Follow this Equipment

The only simultaneous absorbance and fluorescence system for water quality analysis!. Aqualog now offers new deep UV absorbance and fluorescence excit

The only simultaneous absorbance and fluorescence system for water quality analysis!. Aqualog now offers new deep UV absorbance and fluorescence excitation from 200 nm, in addition to a fully automated sample queue tool to facilitate continuous data acquisition, analysis and export for immediate multivariate analysis with the Solotm package from our partner, Eigenvector Research, Inc. 
 
Aqualog is the only instrument to simultaneously measure both absorbance spectra and fluorescence Excitation-Emission Matrices. EEMs are acquired up to a 100 times faster than with other instruments. Dedicated software automates traceable Quinine Sulfate Unit calibration and correction of inner-filter effects and Rayleigh and Raman scattering lines, enabling rapid export to modeling algorithms like PARAFAC. 
 
multivariate analysis - AqualogAqualog now offers new deep UV absorbance and fluorescence excitation from 200 nm, in addition to a fully automated sample queue tool to facilitate continuous data acquisition, analysis and export for immediate multivariate analysis with the Solotm package from our partner, Eigenvector Research, Inc. 
 
The NEW Aqualog® Datastream Dashboard facilitates completely automated analysis and reporting of a wide range of organic matter parameters that are critical for managing and optimizing the drinking water treatment process. The Datastream parameters have been selected to specifically target disinfection by-product issues, algal issues and other contamination components
Active Questions & AnswersAsk a Question

There are no current Discussions

Fluorescence Detector Service ProvidersView All (21)

Request Support
Features of Aqualog®
  • The only true simultaneous absorbance-fluorescence system available * TE-cooled CCD fluorescence emission detector for rapid data acquisition up to 100 times faster than any other benchtop fluorometer * Corrected UV-VIS absorbance detection path for stability and accuracy * Double grating excitation monochromator for superior stray light rejection * Matching bandpass for absorbance and fluorescence spectra * Automatic sample changer option (2 or 4 position) * Compatible with flow cells and titrator * Optimized experiment set-up menus minimize user confi guration time * Complete NIST-traceable corrected fluorescence spectra automatcally generated * Spectral and kinetic analysis tools for both absorbance and fluorescence data * Methods and batch protocols for automating multiple sample measurement
General Specifications

There are no General Specifications available.

Additional Specifications

Choice of Light source 
Standard: 150W ozone-free vertically mounted xenon arc lamp 
Extended-UV: 150W vertically mounted xenon arc lamp 
 
Excitation range
 
230 nm to upper limit of emission detector 
200 nm to upper limit of emission detector 
 
Excitation bandpass 
5 nm 
 
Excitation monochromator 
Subtractive double monochromator 
 
Excitation gratings 
1200 gr/mm, 250 nm blaze 
 
Excitation wavelength accuracy 
±1 nm 
 
Choice of Detector 
UV-Visable 
Red-extended 
 
Emission range 
250-620 nm 
250-800 nm 
 
Emission grating 
405 gr/mm; 250 nm blaze 
285 gr/mm; 350 nm blaze 
 
Hardware pixel binning
 
0.41, 0.82, 1.64, 3.28 nm/pixel 
0.58, 1.16, 2.32, 4.64 nm/pixel 
 
Emission bandpass 
5 nm 
 
Emission spectrograph 
Fixed, aberration-corrected 140 mm focal length 
 
Emission detector
 
TE-cooled back-illuminated CCD 
 
Emission integration time 
5 ms minimum 
 
CCD gain options 
2.25 e-/cts in high gain, 4.5 e-/cts in medium gain, 9 e-/cts in low gain 
 
Sensitivity 
Water-Raman SNR > 20,000:1 (RMS method) (350 nm excitation, 30s integration) 
 
Weight 
32.72 kg (72 lbs) 
 
Dimensions 
LWH (618 x 435 x 336 mm); (24" x 17" x 13") 
 
Scanning range 
200 - 800 nm (UV lamp) 
230-800 nm (Standard lamp) 
 
Bandpass 
5 nm 
 
Slew speed 
Max 500 nm/s 
 
Optical system 
Corrected single-beam 
 
Detector 
Si photodiode 
 
Wavelength accuracy 
±1 nm 
 
Wavelength repeatability 
+/- 0.5 nm 
 
Photometric accuracy
 
±0.01 AU from 0 to 2A 
 
Photometric stability 
<0.002 AU per h 
 
Photometric repeatability 
+/- 0.002 AU (0 to 1 AU) 
 
Stray light 
<1% measured with KI standard