Lumex Instruments - Atomic Absorption Spectrometer MGA-1000
Manufactured by Lumex Instruments
The MGA-1000 series of graphite furnace atomic absorption spectrometers crystallizes intuitive technology and...
The MGA-1000 series of graphite furnace atomic absorption spectrometers crystallizes intuitive technology and exceptional handling. The unique High Frequency Zeeman background correction system ensures effective compensation of intense and rapidly changing spectral interferences and brings routine analysis of complex samples to new heights.
KEY TECHNOLOGIES
Zeeman High Frequency Polarization Modulation (ZHFPM) background correction
Simultaneous measurement of total and background absorptions provides accurate background correction even for complex sample matrices
High intensity electrodeless discharge lamps (EDL) ensure lower detection limits compared to traditional hollow cathode lamps (HCL) and make possible direct As and Se determination at ppb and sub-ppb levels without hydride generation technique
The stabilized temperature platform furnace (STPF) concept realized
Instrument drifts arising from various sources e.g. spectral lamps, electronics, are effectively compensated by quasi-double beam optical system. No warm-up needed
Cold Vapor / Hydride Generation technique available
Noiseless operation and small size
INDUSTRIES
Graphite furnace atomic absorption spectroscopy (GFAAS) is an excellent solution for trace element analysis in various fields:
Environmental monitoring
Water treatment
Food and Beverages
Agriculture
Pharmaceutics and Life science
Medicine and Biochemistry
Forensics and Clinical research
Petrochemistry and Refineries
Process control and Material analysis
Science and Research
For laboratories with low-moderate sample throughput, GFAAS secures optimal combination of high selectivity, sensitivity and low-cost analysis.
HIGH ANALYTICAL PERFORMANCE
Simple and direct analysis of samples with high salt and organic content including blood and biological fluids, sea water and brines, some food and soil digests, etc.
Lower detection limits for the elements for which EDLs are available. Thus Sb, As, Bi, Cd, Pb, P, Se, Te, Tl, Sn and Zn can be determined in variety of complex samples without pre-concentration and hydride generation technique
AUTOMATION & USABILITY
Plug & Measure convenience
User-friendly software for setting operating conditions, results storage and processing
PC controlled 6-lamp turret designed for main types of Hollow Cathode Lamps and Electrodeless Discharge Lamps
Smart autosampler with all routine functions
Safety self-control
LOW COST ANALYSIS
Standard graphite tubes (28 mm) of both Massman and L’vov type are available
The autosampler accommodates standard plastic vials and Eppendorf®-type tubes
Prolonged tube lifetime
KEY TECHNOLOGIES
Graphite furnace atomic absorption spectroscopy (GFAAS) is an excellent solution for trace element analysis in various fields:
For laboratories with low-moderate sample throughput, GFAAS secures optimal combination of high selectivity, sensitivity and low-cost analysis.
Active Questions & AnswersAsk a Question
There are no current Discussions
Need Equipment Support?
Documents & ManualsView All Documents
Features
There are no Features available.
General Specifications
Depth | 475 cm |
Spectral Range | 185 to 900 nm |
Height | 310 cm |
Width | 800 cm |
Maximum Temperature | 3000 °C |
Weight | 50 kg |
Additional Specifications
Furnace heating rate up:
up to 4500 K/sec
Argon flow rate (high purity argon):
< 1 l/min
Tube lifetime:
800 heating cycles