Beckman Coulter - MET ONE 6000
Remote Air Particle Counters
Active Questions & AnswersAsk a Question
There are no current Discussions
Particle Counters Service ProvidersView All (3)
Documents & Manuals
There are no Documents or Manuals available.
Features of MET ONE 6000
Industry leading particle counting on-line reliability ISO 21501 compliant with built-in flow sensor for CFR21 data accuracy and reliability.
Flexible for all your remote air particle counter applications Flexible particle count size in 0.3um or 0.5um sensitivity at 0.1cfm or 1.0cfm flow rates.
Industry leading quality and reliability Long life laser diode for guaranteed continuous monitoring.
Easy to integrate with any facility monitoring system Open architecture design to communicate seamlessly with your system Multiple communication options: Pulse, Serial Modbus, Ethernet, Wireless and Analog.
Faster trouble shooting and reduced downtime MET ONE 6000 remote air particle counters can be diagnosed for flow, sensor, or communication failures to provide advanced notification of hardware problems Count alarm and alerts enable visual notification via built-in status lights.
Easy installation options – reducing cost Reduce initial installation cost with MET ONE 6000 remote air particle counter’s integrated multi-parameter measurements. Optional relative humidity and temperature probe integrates RH/T data with particle count data in same data stream.
General Specifications
Depth | 8.13 cm |
Height | 10.87 cm |
Width | 13.56 cm |
Weight | 0.68 kg |
Additional Specifications
Number of Size Channels
Standard 2 or Optional 4 (pulse units have 2 channels only)
Light Source
Long Life Laser™ diode
Enclosure
304 stainless steel
Power Requirements
9 to 28 VDC
Power Consumption
Serial and pulse unit: 3.3 W maximum
Ethernet unit: 4.3 W maximum
Operating Temperature
10 to 32 °C (50 to 90 °F), 5 to 95% relative humidity, non-condensing
Storage Temperature
-40 to 70 °C (-40 to 158 °F), 5 to 98% relative humidity, non-condensing
Signal Output Options
Pulse, Analog 4–20 mA, Serial RS-232, Serial RS-485 with Modbus RTU or FXB communication protocol, Ethernet with TCP/IP protocol
Vacuum Requirements
At least 40.6 cm (16 inches) Hg
Flow Control
Through critical orifice
Coincidence Loss
Model 6002/6003: 5% at 2,000,000 particles/ft3, (70,600,000 particles/m3)
Model 6015: (except Pulse) 5% at 400, 000 particle/ft3, (14,000,000 particles/m3)
False Count Rate
One or less in five minutes
Status Indicator
Multi-colored LED for normal status, count alarm, count alert, sensor failure, flow failure or communication failure