Malvern Panalytical - Zetasizer AT

Manufactured by  Malvern Panalytical
Follow this Equipment

On-line measurement of particle size using Dynamic Light Scattering

The Zetasizer AT combines Malvern's industry leading Dynamic Light Scattering (DLS) technology with two decades of expertise in design and automation of process interfaces for analytical instrumentation.

The Zetasizer AT is a fully-automated nanoparticle characterization system, delivering lab quality results on-line and in a timely manner for process optimization.

Active Questions & AnswersAsk a Question

There are no current Discussions

Particle Size Analyzers Service ProvidersView All (2)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Zetasizer AT
  • Ability to monitor processes involving nanoscale particles or droplets in real time e.g. homogenization
  • Ability to track changes in particle size with time and determine when the end point is reached
  • Automated sampling system facilitates sample extraction, degassing and dilution to the required concentration minimizing operator input
  • Sampling system designed to minimize sample consumption
  • Easy transfer of laboratory DLS methods to on-line system
  • Utilizes Malvern's Universal Interface for system integrators (Malvern Link II)
General Specifications
Depth 66 cm
Height95 cm
Width 98 cm
Weight120.5 kg
Additional Specifications

Measurement range 
1nm  1 micron* (diameter) (0.3nm  10 microns* at-line) 
Measurement principle 
Dynamic Light Scattering 
Accuracy 
Better than +/-2% on NIST traceable latex standards 
Precision / Repeatability 
Better than +/-2% on NIST traceable latex standards 
Sensitivity 
0.1mg/mL (Lysozyme) 
Sample Interface 
1/16" OD (0.040" ID) or 1/8" OD (1/16" ID) tube 
Pressure 
0 - 3 Bar (0- 44 psi) at sample inlet. 
Temperature 
0  90 °C (flow cell only) 
Minimum sample volume 
4ml (application dependent) 
Dilution 
up to 1:50 / infinite (application dependent) 
Power 
110-240VAC, 6A max, 50-60Hz 
Compressed Air for optics purge (optional) 
2-8 Bar (29-116 psi) per ISO8573-1:2010 [1.7.2] 
Ingress Protection (IP) rating 
IP54

Looking for New or Used Equipment?

Shop on LabX