EPPENDORF - 5418/5418R
Manufactured by EPPENDORF
Routine centrifugation with low-to-moderate sample throughput
Models 5418 and 5418 R are ultra-compact 18-place centrifuges that fits on even the smallest lab bench. About the size of a standard sheet of paper, it provides up to 16,873 x g (14,000 rpm) and offers a broad range of features that make your lab work easier. The centrifuge is whisper quiet—even when running without the rotor lid—to improve your work environment.
The compact Centrifuge 5418 has a capacity for 18 x 1,5/2,0 ml tubes and it can achieve a maximum rcf of 16,873 x g/14,000 rpm. This Micro Centrifuge is exclusively equipped with an aerosol tight rotor, which can be operated without the rotor lid for customer convenience but without any significant increase in noise level. The Centrifuge 5418 is perfect for customers that demand a high quality centrifuge for processing a reduced amount of samples.
The compact Centrifuge 5418 has a capacity for 18 x 1,5/2,0 ml tubes and it can achieve a maximum rcf of 16,873 x g/14,000 rpm. This Micro Centrifuge is exclusively equipped with an aerosol tight rotor, which can be operated without the rotor lid for customer convenience but without any significant increase in noise level. The Centrifuge 5418 is perfect for customers that demand a high quality centrifuge for processing a reduced amount of samples.
Active Questions & AnswersAsk a Question
Centrifuge Service ProvidersView All (22)
Documents & Manuals
There are no Documents or Manuals available.
Features of 5418/5418R
- Compact footprint to fit on crowded lab benches
- QuickLock™ rotor lid closure for fast and reliable lid locking
- Temperature range from 0°C to 40°C
- OptiBowl® design allows for whisper quiet operation without rotor lid to reduce handling time and benefit your work environment
- FastTemp function for fast pre-cooling and maximum temperature accuracy inside the rotor
- Maintains constant 4°C at maximum speed
- Built-in condensation drain to eliminate water accumulation and prevent corrosion
- ECO shut-off engages after 8 hours of nonuse to reduce energy consumption and extends compressor life (can be deactivated). Saves you up to 47% energy overnight.
- 18 place rotor
- Max. 16,873 x g
- Remarkably silent, even without rotor lid
- Standard, certified aerosol-tight rotor
- Convenient Eppendorf two-knob operation
- rpm/rcf converter
- Short-spin
- Automatic lid opening following application sequence
- Very small footprint
- Intended Use (pdf-file)
General Specifications
There are no General Specifications available.
Additional Specifications
Max. kinetic energy: 2550 Nm
Permissible density of material to be centrifuged: 1.2 g/ml
Rel. humidity: 10 - 75 %
Setup height max.: 2000 m above NSL
Startup time: (230 V) 15 s
Deceleration time: (230 V) 15 s
Startup time: (120 V) 15 s
Deceleration time: (120 V) 15 s
Startup time: (100 V) 18 s
Deceleration time: (100 V) 15 s
Fuses: 2.5 AT (230 V)
5.0 AT (120 V / 100 V)
Overvoltage category II
Degree of contamination: 2
Series Modals
EPPENDORF - 5418
[+] View Model SpecificationsDepth | 30 cm |
Height | 21 cm |
Width | 21 cm |
Orientation | Benchtop |
Peak Power | 170 W |
Noise Level | 51 dB |
Weight | 17 lbs |
Speed Range | 100 to 14000 rpm |
Centrifuge Type | Microcentrifuge |
Maximum Capacity | 18 |
Maximum Speed | 14000 rpm |
Refrigerated | No_ |
Additional Model Specifications | Max g-force: 16,873 rcf Number of rotors: 1 Capacity for 1.5 mL tubes: 18 Acceleration time to max speed: 13 s (in FA-45-18-11) Braking time from max speed: 13 s (in FA-45-18-11) Acceleration/Brake speed settings: 1 |
EPPENDORF - 5418R
[+] View Model SpecificationsDepth | 460 mm |
Height | 250 mm |
Width | 300 mm |
Orientation | Benchtop |
Peak Power | 320 W |
Noise Level | 57 dB |
Weight | 48.5 lbs |
Speed Range | 100 to 14000 rpm |
Centrifuge Type | Microcentrifuge |
Maximum Capacity | 18 |
Maximum Speed | 14000 rpm |
Refrigerated | Yes_ |
Additional Model Specifications | Refrigerated Max g-force: 16,873 rcf Number of rotors: 1 Capacity for 1.5 mL tubes: 18 Acceleration time to max speed: 13 s (in FA-45-18-11) Braking time from max speed: 13 s (in FA-45-18-11) Acceleration/Brake speed settings: 1 |