HORIBA - EMIA-Pro Carbon/Sulfur Analyzer
Enables everyone to become a professional analyst
The EMIA-Pro, Carbon/Sulfur Analyzer is based on HORIBA’s vast knowledge of Non-Dispersive Infrared (NDIR) technology and our wealth of experience acquired from our EMIA series.
EMIA-Pro offers improved cleaning efficiency, enhanced user-friendly software, durability, operability and maintenance to ensure efficient measurements and its shortened measurement cycle time accelerates repetitive measurements up to 200 times, thus improving development and manufacturing speed.
Active Questions & AnswersAsk a Question
There are no current Discussions
Elemental Analyzers Service ProvidersView All (6)
Documents & ManualsView All Documents
Features of EMIA-Pro Carbon/Sulfur Analyzer
Reduced Maintenance Time
Maintenance time has been reduced by 50%, to 30 minutes with our reduction in the number of components and smart mechanical design.
Increased Measurement Efficiency
One measurement cycle, “start analysis-burn sample-display result-cleaning”, takes just 70 seconds! This allows more than 50 measurements per hour.
Long-term Durability
The unique newly-developed cleaning mechanism (patent pending) clears the dust on the filter, allowing the operator to continue to use the EMIA-Pro without maintenance cleaning for up to 200 measurements (depending on sample and measurement conditions)*.
Advanced Operation software
A critical focus was to ensure that the EMIA-Pro was “easy-to-use”. In particular, we focused on the layout, operating menus and functions. We have included a self-diagnostic monitor checking capability to check the status of the connected device, an alarm function and we have expanded 3 navigations to facilitate daily operation.
- Analysis Navigation: This function proposes best measurement condition based on our experience and know-how, and sets up the measurement conditions.
- Troubleshooting navigation: Automatic diagnosis system and troubleshooting navigation.
- Maintenance Navigation: Periodic preventive maintenance work navigations by photos and movies.
Upgradable
The customer can extend the system’s functionality at any time, even after installation, depending upon their needs. Upgrades include:
- Carrier gas purifier for ultra-low concentration carbon measurement
- Auto-sampler (20 positions)
- Halogen trap unit
General Specifications
Depth | 724 mm |
Height | 710 mm |
Width | 499 mm |
Weight | 134 kg |
Additional Specifications
Required sample amount
1g
Typical Meas. time
70 seconds/cycle
Carbon
Meas. Range (m/m): 1.6 ppm - 6.0%
Blank accuracy: σ ≤ 0.8 ppm
Sulfur
Meas. Range (m/m): 2.0 ppm - 1.0%
Blank accuracy: σ ≤ 1 ppm
Sample meas. Accuracy
Carbon: SD ≤ 2.0 ppm or RSD ≤ 1.0%
Sulfur: SD ≤ 2.0 ppm or RSD ≤ 1.5%
Utility
Carrier gas: Oxygen (Purity; 99.5%, Pressure; 0.3MPa)
Operation gas: Nitrogen (Purity; 99.5%, Pressure; 0.35MPa)
Power
200/220/240V, 50/60Hz, 5kVA
Data processing and operation
USB data communication with PC Windows 8.1
Touch panel or Key board and mouse operation