HORIBA - SLFA-2100/2800 X-ray Fluorescence Sulfur-in-Oil Analyzers
Manufactured by HORIBA
The SLFA-2100/2800 are designed specifically to meet the recent demanding needs of measuring the new low sulfur fuels,...
The SLFA-2100/2800 are designed specifically to meet the recent demanding needs of measuring the new low sulfur fuels, diesel and RFG.
Using the X-ray fluorescence technique, fast and accurate measurements can be carried out in compliance with the ASTM D4294 method, either in the lab or in the field.
A detection limit of 5 ppm sulfur takes you to new realms of measurement sensitivity. HORIBA's high sensitivity analyzer makes quick, accurate measurements of petroleum fuel samples from very low to high sulfur concentrations - i.e., all the way from gasoline to heavy fuel oils. A large, easy-to-read LCD screen, simple function keys, alphanumeric keypad and other features make the SLFA-2100/2800 Series so easy to use.
The SLFA-2800 has an automated turntable to measure up to eight samples unattended, while the SLFA-2100 is a single-sample version.
Using the X-ray fluorescence technique, fast and accurate measurements can be carried out in compliance with the ASTM D4294 method, either in the lab or in the field.
A detection limit of 5 ppm sulfur takes you to new realms of measurement sensitivity. HORIBA's high sensitivity analyzer makes quick, accurate measurements of petroleum fuel samples from very low to high sulfur concentrations - i.e., all the way from gasoline to heavy fuel oils. A large, easy-to-read LCD screen, simple function keys, alphanumeric keypad and other features make the SLFA-2100/2800 Series so easy to use.
The SLFA-2800 has an automated turntable to measure up to eight samples unattended, while the SLFA-2100 is a single-sample version.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked bykcprus10
Need Equipment Support?
Documents & Manuals
There are no Documents or Manuals available.
Features of SLFA-2100/2800 X-ray Fluorescence Sulfur-in-Oil Analyzers
- Wide measurement range (0-9.999%) is acceptable for very small content as in diesel oil or high sulfur content as in heavy oil. * Calibration graphs have been extended and the analyzer selects automatically out of seven scales (0.002/0.01/0.05/…/10%). * High precision measurement (Repeatability:1.6ppm, Lower detection limit: 5ppm) Due to the miniature X-ray tube, extremely precise results can be achieved without purging of He or another gas and this system also include a correction function for temperature and atmospheric pressure without He purge. * SLFA-2800 is equipped with a turntable, which makes sequential measuring of up to eight samples. * Standards applied to device measurement functions: JIS K2541, JIS B7995(Japan), ASTM D4294(USA), ISO 8754
General Specifications
Depth | 19.7 in |
Height | 9.1 in |
Width | 19.9 in |
Power Requirements | 100 V AC to 240 V±10%, 50/60 Hz |
Peak Power | 150 VA |
Additional Specifications
Principle:
X-ray fluorescent analysis
Sample:
Sulfur content in petroleum products, such as heavy oil, naphtha, crude oil, etc.
Element:
Sulfur
Range:
0-9,999%
Detection Limit:
5 ppm or less
Accuracy:
Less than 5 ppm (with a sample containing 1% sulfur)
Less than 1.6 ppm (with a sample containing 0% sulfur)
Calibration:
Calibration made using any standard sample (2 to 20 calibration points)
Sample requirement:
4-10 ml
Measurement time:
User set from 10 to 600 seconds
External output:
Printout of results and digital data output(RS-232C interface)