DESCRIPTION
Atomic Absorption Spectrometer. Single element analyser requiring changeing to individual lamps for elemental analysis.

The PerkinElmer AAnalyst™ 800 high-performance atomic absorption spectrometer with WinLab32 for AA software features an automated motorized atomizer exchange that allows switching between flame and graphite furnace AA by a simple software command. The instrument is equipped with a high performance burner system , TotalFlow™ gas controls for flame AA and a Transversely Heated Graphite Furnace (THGA) with longitudinal Zeeman-effect background corrector.

FEATURES

The graphite furnace system includes True Temperature Control (TTC) and integrated platform tubes providing full Stabilized Temperature Platform Furnace (STPF) conditions for almost interference-free trace metal analysis. The built-in, high-capacity AS-800 autosampler adds exceptional productivity and flexibility.        The AAnalyst 800 utilizes the industry leading WinLab 32 for AA software that brings a new level of productivity to high-performance atomic absorption. The optional Enhanced Security (ES) software adds additional capabilities needed for regulatory requirements such as 21 CFR Part 11.

SPECIFICATIONS
Graphite Furnace1
Lamps8-lamp holder
Maximum Temperature2600 °C in steps of 10 °C
Zeeman1
Wavelength Range190 to 900 nm
Power Requirements 230 V (±10%), 50/60 Hz (±1%),
Peak Power5000 VA (max)
Weight187 kg (without controller and cooling system)
Advertisement
QUESTIONS & ANSWERS

0

Answers
-Sadi
9 days ago
Unable to evaluate QC
I am suddenly getting this message after my QC runs! Can anyone help me understand? Furnace.

Reply

0

Answers
-Becky W
2 months ago
Problem with autosampler AS800- please help
Error Message: Furnace autosampler has timed out. The furnace autosampler was unable to cycle normally and is now in standby mode



During a stability check, the calibration, blanks, and LFB came out fine and then we were prompted with this error message.

The autosampler tray is now not moving during a run and therefore the sampling arm is improperly placed. The tray is rotating fine but does not move left and right. During the rinse phase, the tip is injecting into a sample cup.



Thank you for your help

Reply

1

Answer
-bhupi80singh
3 months ago
Purchase
Hi, I am planning to buy Perkin ELmer 800 AAS. I am wondering what are other make and models that I should consider to make my final decision?

Thanks,

Bhupinder

Reply