PerkinElmer - AAS 800
Manufactured by PerkinElmer
Atomic Absorption Spectrometer. Single element analyser requiring changeing to individual lamps for elemental analysis.
The PerkinElmer AAnalyst™ 800 high-performance atomic absorption spectrometer with WinLab32 for AA software features an automated motorized atomizer exchange that allows switching between flame and graphite furnace AA by a simple software command. The instrument is equipped with a high performance burner system , TotalFlow™ gas controls for flame AA and a Transversely Heated Graphite Furnace (THGA) with longitudinal Zeeman-effect background corrector.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked bybyrmakgln3426
Need Equipment Support?
Documents & ManualsView All Documents
Features of AAS 800
The graphite furnace system includes True Temperature Control (TTC) and integrated platform tubes providing full Stabilized Temperature Platform Furnace (STPF) conditions for almost interference-free trace metal analysis. The built-in, high-capacity AS-800 autosampler adds exceptional productivity and flexibility.
The AAnalyst 800 utilizes the industry leading WinLab 32 for AA software that brings a new level of productivity to high-performance atomic absorption. The optional Enhanced Security (ES) software adds additional capabilities needed for regulatory requirements such as 21 CFR Part 11.
General Specifications
Graphite Furnace | 1 |
Lamps | 8-lamp holder |
Maximum Temperature | 2600 °C in steps of 10 °C |
Power Requirements | 230 V (±10%), 50/60 Hz (±1%), |
Peak Power | 5000 VA (max) |
Weight | 187 kg (without controller and cooling system) |
Zeeman | 1 |
Wavelength Range | 190 to 900 nm |