BioTek - Epoch Multi-Volume
Manufactured by BioTek
Epoch Multi-Volume Spectrophotometer System
Epoch™ Microplate Spectrophotometer is coupled with an innovative device like the new Take3™ Multi-Volume Plate, the result is a system that’s designed to allow the scientist to run a variety of assays, in a variety of sample vessels, and volume ranges.
Active Questions & AnswersAsk a Question
There are no current Discussions
Microplate Readers Service ProvidersView All (12)
Documents & ManualsView All Documents
Features of Epoch Multi-Volume
As a system, Epoch and Take3 are ready to take on the challenges of today’s researchers:
- Save valuable sample by using Take3 to quantitate 2 µL volumes of DNA, RNA and proteins
- Monitor cell growth using the Epoch’s well scanning capability
- Run spectral scans of 2 µL samples, or samples contained in a microplate, standard cuvette or the patented BioCell quartz Cuvette
- Dial in individual wavelengths from 200 – 999 nm Perform rapid qualitative or quantitative analysis in microplates from 6- to 384-wells
- Measure kinetic rates of reaction
- Obtain quick 260/280 ratios in 2 µL samples or in a microplate
- Validate the miniaturization of your assay…from a full cuvette to a microplate to a microvolume…on one system
- Run multiplex assays quickly and easily
- Save space on the benchtop with its very small footprint - Avoid costly maintenance with the robust, thoughtfully engineered optics and hardware
- Never purchase costly absorbance filters
All of these challenges and more are addressed with the Epoch Multi-Volume Spectrophotometer system…at a price that doesn’t challenge the budget. The system includes the Epoch Microplate Spectrophotometer with Gen5 Software and the Take3 Multi-Volume Plate with its own preprogrammed Gen5 module.
General Specifications
Bandwidth | 5 nm |
Depth | 12.5 in |
Height | 7.7 in |
Width | 12 in |
Sample Formats | 6-, 12-, 24-, 48-, 96- and 384-well plates |
Weight | 14 lbs |
Read Method | Endpoint, kinetic, spectral scanning, well area scanning |
Wavelength Range | 200 to 999 nm |
Scanning Speed | 96-well Normal: 49 seconds. 96-well Rapid: 38 seconds. 96-well Sweep: 15 seconds. |
Additional Specifications
Wavelength Accuracy: ±2 nm
Wavelength Repeatability: ±0.2 nm
Absorbance Range: 0 to 4.0 OD
Absorbance Resolution: 0.0001
Accuracy:
0.0 to 2.0 OD: ±1% ±0.010 OD
2.0 to 2.5 OD: ±3% ±0.010 OD
Repeatability:
0.0 to 2.0 OD: ±1% ±0.005 OD
2.0 to 2.5 OD: ±3% ±0.005 OD
Linearity:
0.0 to 2.0 OD: ±1% ±0.010 OD
2.0 to 2.5 OD: ±3% ±0.010 OD