Shimadzu - GC-8A
Manufactured by Shimadzu
Dedicated Single Detector GC
The GC 8 is a single inlet, single detector, gas chromatograph. It's very simple yet robust design makes it perfect for a teaching lab or for a dedicated application in a QC environment. Its all metal construction provides for a GC that is very rugged and can easily withstand the rigors of an industrial QC setting. This system offers a choice of three packed column detectors: FID, ECD or TCD.
Active Questions & AnswersAsk a Question
There are no current Discussions
GC Systems Service ProvidersView All (42)
Documents & ManualsView All Documents
Features of GC-8A
- Simple design with emphasis on functionality
- Two-stage overheat prevention for protection of columns
- Temperature control offers excellent stability and response
- Small Footprint with large capacity, cubic oven
General Specifications
There are no General Specifications available.
Series Modals
Shimadzu - GC-8AIE
[+] View Model SpecificationsDepth | 405 mm |
Height | 435 mm |
Width | 440 mm |
Power Requirements | AC100/115V or 220V as ordered. 1500VA max. 50/60Hz. |
Weight | 26 kg |
Additional Model Specifications | Column Oven Temperature range: Room temperature to 399°C Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Temperature Readout (Monitor): When the temperatures of the column oven and the injection port/detector have reached the set points, the respective READY lamps are lit. Heating speed: From room temperature to 350°C in 13 minutes. Cooling speed: From 350°C to 100°C in 3.5 minutes. Column: Stainless steel: 12m x 1 Glass: 5.4m x 1 Overheat Protection: Two-stage protection: (1) If the temperature exceeds the programmed (preset) temperature by 50°C, the heater will be automatically turned off. (2) If the temperature reaches about 420°C, the heater will be automatically turned off. Temperature of Injection Port/Detector Temperature range: Room temperature to 350°C (10°C steps) Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Temperature Readout (Monitor): When the temperatures of the column oven and the injection port/detector have reached the set points, the respective READY lamps are lit. Overheat Protection: If the temperature reaches about 350°C, the heater will be automatically turned off. Injection Port: On-column injection ports Flow Control Carrier gas: One pressure regulator One column inlet pressure gauge Electron Capture Detector (ECD) Detector: Coaxial type cell and purge flow line Radiation source: 63Ni 370MBq Detection limit: 0.2pg/s for Υ-BHC Dynamic range: 104 Maximum temperature: 350°C Standing current: 0.5, 1.0, 2.0mA Other: Column oven heating starts after the injection port / detector reaches the set temperature. |
Shimadzu - GC-8AIF
[+] View Model SpecificationsDepth | 405 mm |
Height | 570 mm |
Width | 440 mm |
Power Requirements | AC100/115V or 220V as ordered. 1500VA max. 50/60Hz. |
Weight | 34.5 kg |
Additional Model Specifications | Column Oven Temperature range: Room temperature to 399°C Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Heating speed: From room temperature to 350°C in 13 minutes. Cooling speed: From 350°C to 100°C in 3.5 minutes. Column: Stainless steel: 6m x 2 Glass: 5.4m x 2 Overheat Protection: Two-stage protection: (1) If the temperature exceeds the programmed (preset) temperature by 30°C, the heater will be automatically turned off. (2) If the temperature reaches about 420°C, the heater will be automatically turned off. Temperature of Injection Port/Detector Temperature range: Room temperature to 400°C (10°C steps) Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Overheat Protection: If the temperature reaches about 420°C, the heater will be automatically turned off. Injection Port: On-column injection ports Temperature Readout The three temperatures (injection port/detector, column oven, and auxiliary) are selectively indicated on the pyrometer, the selection being made by pushbuttons. Flow Control Carrier gas: Two pressure regulators Two column inlet pressure gauges Hydrogen: Two pressure regulators Two pressure gauges Air: One pressure regulator One pressure gauge Flame Ionization Detector (FID) Detector: Dual cell with cylindrical collectors Maximum temperature: 400°C Detection limit: 3 x 10-12 g/s for diphenyl Dynamic range: 107 Jet: Mode of quartz Electrometer Sensitivity: 1 x 10-12 A/mV Noise: Below 10-14 A Drift: Below 2 x 10-14 A/h (Constant room temperature) Input attenuation: 1, 10, 102, 103, 104 Output attenuation: 1 to 1024 (binary step) and ? Linearity: Wider than 105 (fixed input attenuation) Background suppression: ±6.4 x 10-10A (NORM mode) Signal output: 0 to 1mV for recorder 0 to 1V for data processor |
Shimadzu - GC-8AIT
[+] View Model SpecificationsDepth | 405 mm |
Height | 435 mm |
Width | 440 mm |
Power Requirements | AC100/115V or 220V as ordered. 1500VA max. 50/60Hz. |
Weight | 26.5 kg |
Additional Model Specifications | Column Oven Temperature range: Room temperature to 399°C Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Temperature Readout (Monitor): When the temperatures of the column oven and the injection port/detector have reached the set points, the respective READY lamps are lit. Heating speed: From room temperature to 350°C in 13 minutes. Cooling speed: From 350°C to 100°C in 3.5 minutes. Column: Stainless steel: 6m x 2 Glass: 5.4m x 2 Overheat Protection: Two-stage protection: (1) If the temperature exceeds the programmed (preset) temperature by 50°C, the heater will be automatically turned off. (2) If the temperature reaches about 420°C, the heater will be automatically turned off. Temperature of Injection Port/Detector Temperature range: Room temperature to 400°C (10°C steps) Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Temperature Readout (Monitor): When the temperatures of the column oven and the injection port/detector have reached the set points, the respective READY lamps are lit. Overheat Protection: If the temperature reaches about 420°C, the heater will be automatically turned off. Injection Port: On-column injection ports Flow Control Carrier gas: Two pressure regulators Two column inlet pressure gauges Thermal Conductivity Detector (TCD) Detector: Differential type of semi-diffusion flow system Four rhenium-tungsten filaments (100Ω each) Maximum temperature: 400°C Bridge current: Constant current system (OFF, 60, 70, 80, 90, 100, 120, 140, 160, 180, 200mA) Filament protection circuit: Incorporated Others: (1) Column oven heating starts after the injection port / detector reaches the set temperature. (2) Cannot be used with a preparative system. |
Shimadzu - GC-8APFp
[+] View Model SpecificationsDepth | 405 mm |
Height | 650 mm |
Width | 440 mm |
Power Requirements | AC100/115V or 220V as ordered. 1500VA max. 50/60Hz. |
Weight | 35 kg |
Additional Model Specifications | Column Oven Temperature range: -100°C to 400°C (The optional cryogenic work attachment and cryogenic valve unit are necessary for sub-ambient temperature operation.) Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Heating speed: From room temperature to 350°C in 13 minutes. Cooling speed: From 350°C to 100°C in 3.5 minutes. Column: Stainless steel: 12m x 1 Glass: 5.4m x 1 Overheat Protection: Two-stage protection: (1) If the temperature exceeds the programmed (preset) temperature by 30°C, the heater will be automatically turned off. (2) If the temperature reaches about 420°C, the heater will be automatically turned off. Temperature Programmer/Timer Initial temperature: -100°C to 399°C (1°C steps) Final temperature: 0°C to 400°C (10°C steps) Programming rate: 0.5, 1, 2, 3, 4, 5, 6, 8, 10, 16, 20, 32°C/min. Temperature of Injection Port/Detector Temperature range: Room temperature to 400°C (10°C steps) Temperature control: Proportional-integration type of zero cross switching system Temperature control accuracy: ±0.1°C Overheat Protection: If the temperature reaches about 420°C, the heater will be automatically turned off. Injection Port: On-column injection ports Temperature Readout: The three temperatures (column oven, injection port/detector, and auxiliary) are selectively indicated on the pyrometer, the selection being made by pushbuttons. Flow Control Carrier gas: One primary pressure regulator One primary column inlet pressure gauge One differential flow controller One column inlet pressure gauge Hydrogen: One pressure regulator One pressure gauge Air: One pressure regulator One pressure gauge Flame Photometric Detector (FPD) Selective detection of sulfur and phosphorus (by filter interchange) Optional filter: 394nm (S), 526nm (P) Detection limit: 5 x 10-11 gS/s for S in thiophene 1.4 x 10-12 gP/s for P in DDVP or parathion Power supply: -700V, stabilized Maximum temperature: 350°C (air cooling) Electrometer Sensitivity: 1 x 10-10 A/mV Noise: Below 10-12 A Drift: Below 2 x 10-12 A/h (Constant room temperature) Input attenuation: 1, 10, 102, 103 Output attenuation: 1 to 1024 (binary step) and ? Linearity: Wider than 105 (fixed input attenuation) Background suppression: ±6.4 x 10-8A Signal output: 0 to 1mV for recorder 0 to 1V for data processor Special accessories: FID monitor and electrometer for monitor (P/N 221-23950-91) Filter for S (P/N 221-00892-01) Filter for P (P/N 221-00897-01) |