Pharmatest USA - PTWS 800D
Compact 8-Station Dissolution Testing Instrument
The PTWS 800D features the 3CS three-point individual centering system for each dissolution vessel. The vessels are held in position by three adjustable noses and are inserted into the instrument support framework. The access points for sampling as well as the openings for the tools are contained in an auxiliary, low evaporation, vessel cover.
Active Questions & AnswersAsk a Question
There are no current Discussions
Dissolution Testing Service ProvidersView All (16)
Documents & Manuals
There are no Documents or Manuals available.
Features of PTWS 800D
- Fully USP and EP compliant * 8 stirred positons * 3CS 3-point individual vessel centering system * Centrally located electronic lift drive * Staggered start capability * Options – PT-DL1 data logger * Interfaces - RS232 to connect data logger, I/O port for remote control in offline automated applications
General Specifications
Depth | 45 mm |
Width | 75 cm |
Additional Specifications
Display
LCD, illuminated in different colors
Data Entry
Touch sensitive glass plate, functional touch keys
Acoustic Signal
Programmable acoustic signal for operator information
Timer
Programmable sampling times, wake-up and sleeping mode, operation time information and count-down mode
Printer
External data logger
Number of Stirred Vessels
8 (4 by 2 or 2 by 4 arrangement)
Standard Vessels
1 liter USP/EP Borosilicate glass vessel, each individually coded
Speed Control
25 - 250 RPM
Speed Accuracy
±2% of set speed, typically < 1%
Stirrer Shaft Wobble
Better than 0.2 mm total run out
System Tools
MonoShaft™ stirrer design, USP/EP apparatus 1, 2, 5, 6 tool adapter, cream cell, transdermal patch tools, each tool and vessel individually coded
Heater Range
25 - 45°C
Heater Accuracy
± 0.2°C inside the water bath
Heat Up Process
Energy saving, programmable “wake up” heater function and “sleep mode”
Water Circulation
Water circulated from external heating system through special diffuser inside the water bath
Vibration Elimination
External heating system, spring loaded pump
Calibration
Built-in calibration procedures for speed, temperature control, OQ/PQ interval programmable including alarm indicator
Bench Space Requirment
45 x 75 cm (2 by 4) or 75 x 45 cm (4 by 2)
Certification
All components certified to USP / EP requirements
CE / EMC Certification
All CE / EMC Certification provided