Shimadzu - AA-7000

Manufactured by  Shimadzu
Follow this Equipment

Atomic Absorption Spectroscopy

The AA-7000 series' 3-D optical system is capable of producing maximum performance for both flame detection (Pb 0.015 ppm) and furnace detection (Pb 0.00005 ppm). This is achieved through optimal adjustment of the light beam and digital filter, and by using optical components that restrict light loss.  
  
The optical double-beam photometric system is automatically set for flame measurements, and the high-throughput, single-beam photometric system is automatically set for furnace measurements. With a new graphite furnace design, the AA-7000 Series has achieved improved lower limits of detection for furnace analysis in all fields, as compared to Shimadzu's previous series designs.  
  
The AA-7000 has the smallest footprint of any flame or furnace AA spectrophotometer on the market, combining flame and furnace system mounts in the same compartment. When combined with an optional auto atomizer changer, the AA-7000F creates an integral burner and furnace that remains permanently installed in the combustion chamber. It supports flame analysis of high-concentration samples and furnace analysis of trace samples without changing the atomizer unit.  
  
The AA-7000 series gives users a single high-performance autosampler for both flame and furnace, saving capital cost and set-up time. The ASC-7000 autosampler can layer a solution in the furnace like competitive systems, but takes analysis a step further by truly mixing a sample with four other solutions, such as matrix modifiers, matrix matching, dilution or standard addition samples, for a higher level of automation. Optimal background correction methods are installed as standard, and a high-speed self-reversal method provides correction over a 185 nm to 900 nm range.  
  
Additionally, because safety is an important consideration for flame atomic absorption spectrophotometers, the AA-7000 series are also the first instruments in the world to be equipped standard with a vibration sensor that extinguishes flames automatically. Other safety features include a multimode automatic gas leak check and automatic air-N2O switching for flame-specification instruments, and an argon gas pressure monitor and over-current protection unit for furnace-specification instruments.
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked byaboyosif

Need Equipment Support?

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of AA-7000

Double-Beam Optics and Stable Hardware Achieve Superior Stability   The AA-7000 series’ newly developed 3-D optical system is capable of producing maximum performance for both flame and furnace detection. This is achieved through optimal adjustment of the light beam and digital filter, and by using optical components that restrict light loss.  

Achieves World-Class Lower Limits of Detection   Advances in optics and a new graphite furnace design achieve improved lower limits of detection for furnace analysis (comparison with previous Shimadzu models). This superb analysis performance is possible in all fields.  

Dual-Background Correction Functions   The optimal background correction methods, high-speed self-reversal method (SR method) and deuterium lamp method (D2 method), are installed as standard.   

Advanced Safety Technology   Safety is an extremely important consideration for flame atomic absorption spectrophotometers using acetylene gas. The AA-7000 Series instruments are the first* ones in the world fitted with a vibration sensor as standard. They also incorporate a comprehensive range of safety mechanisms, including gas leak detectors.  

  • Shimadzu in-house investigation. Current as of November 2008.   

Configuration Flexibility   When attached to the AA-7000F, the optional AAC-7000 Auto Atomizer Changer* creates an integral burner and furnace that remains permanently installed in the combustion chamber. It supports both flame analysis of high-concentration samples and furnace analysis of trace samples, without changing the atomizer unit. The integral construction with the atomizer saves space.  

  • The AAC-7000 cannot be attached to an AA-7000G instrument.   

The AA-7000 Series can be upgraded by adding units to allow the system to handle the analysis targets.   

Single Autosampler for Flame/Furnace   The AA-7000 series gives users a single high-performance autosampler for both flame and furnace, saving capital cost and set-up time.

General Specifications
Graphite Furnace1
Lamps6-lamp turret, 2 lamps simultaneously lit
Power Requirements 100, 120, 220, or 230 VAC, 50/60 Hz
Optical SystemDouble-beam
Wavelength Range185 to 900 nm