GC/MSD System

HP 5972 system uses electron impact (EI) ionization and is capable of performing full mass scans or selective ion monitoring (SIM).    
Mass Spectrometer:     
    
Full Scan:     
Mass Range: 1.6-700 amu     
Scan Speed: Up to 1800 amu/sec with 1 amu resolution     
Sensitivity: 10pg of Hexachlorobenzene yields a signal to noise of >10:1     
    
Selective Ion Monitoring:     
    
Mass Range: Up to 50 groups of 30 masses/group     
Sensitivity: 200fg of Hexachlorobezene yields a signal to noise of >10:1
Active Questions & AnswersAsk a Question

Recent Questions & Answers

avatar placemark

Asked bytfr

avatar placemark

Updated byDan Ste.Marie

avatar placemark

Updated byDan Ste.Marie

GC-Single Quad/MSD Service ProvidersView All (1)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of 5972
  • Lower noise by factor of 5X.     
  • Run at lower Emission current 35uA (similar to 5972-5973).     
  • Use 5972/73 filament.     
  • Able to tune in higher repeller voltage (similar to 5972-5973).     
  • Mass range: Extended to 700 AMU