Hewlett Packard 5890 Series II Gas Chromatograph
The HP 5890 Series II provides flexibility in choices among inlets, columns, and detectors through the use of liners and adapters, allowing any standard column to be used without sacrificing performance. Additional flexibility is gained through positions of inlets and detectors relative to each other and through the large internal volume of the oven. There are five inlet options and six detector, as well as a wide variety of flow and pressure control components.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Asked byKiwichemist
GC Systems Service ProvidersView All (42)
Documents & ManualsView All Documents
Features of 5890 Series II
Common Configurations:
- HP 5890 Series II single injector (split/split less or Packed) single FID detector
- HP 5890 Series II single injector (split/split less or Packed) single TCD detector
- HP 5890 Series II dual injectors (split/split less or Packed) dual FID detectors
- HP 5890 Series II dual injectors (split/split less or Packed) dual TCD detectors
- HP 5890 Series II dual injectors (split/split less or Packed) dual FID/TCD detectors
- All of the above systems come with INET communications (HPIB and data system is available.)
General Specifications
Detector | Flame Ionization Detector, Thermal Conductivity Detector |
GC System Type | Modular |
Additional Specifications
HP 5890 Series 2
Heated Devices
- Five heated zones standard: Two detectors, Two inlets, one auxiliary
- Methods stored: Two
Split/Splitless Capillary
- Range to 400°C in 1°C increments
- Back-pressure design permits independent adjustment of split flow rate without
affecting column flow - Septum purge built in at 3 ml/min
- Accepts columns up to 1.2-mm od
- Multimode design includes split and splitless injection
- Splitless purge time variable in 0.01-min elements
Flame Ionization Detector
- Range to 450°C
- Grounded jet and current limited design for operator safety
- Push-button flame ignition
- Minimum detectable: 5 pg carbon/sec, nitrogen carrier at S/N=2
- Linear dynamic range: +/- 10% over a 10 7 range with 0.018-in. id jet
- Conditions: column flow 50 ml/min, 45 ml/min H2, 650 ml/min air, propane
sample
Thermal Conductivity Detector
- Range to 400 °C
- Single-filament (single column) design has fluidic switching of reference and
analytical carrier flows - Minimum detectable :400 pg/ml carrier -- equivalent to 1 pmm neon in 1 ml air
- Linear dynamic range:+/- 5% over 105 range
- Digital gain setting time-programmable through the keyboard Conditions:
detector - 100 °C, 45 ml/min switching and 30 ml/min analytical flow of helium, propane
sample