Beckman Coulter - Allegra X-5
Maximum speed and capacity for high-throughput labs.
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Centrifuge Service ProvidersView All (22)
Documents & Manuals
There are no Documents or Manuals available.
Features of Allegra X-5
Maximize throughput
- Process up to 140 x 13mm blood tubes and 100 x 16mm blood tubes
- Shorten run times with centrifugal forces up to 4,470 x g
- Safely process hazardous samples with an optional BioCertified rotor
Seamlessly integrate with Beckman Coulter analyzers
- Seamlessly integrate with Beckman Coulter analyzers
- Eliminate transfer steps by spinning Beckman Coulter DxC/DxI tube racks directly in the centrifuge
- Run specialized Allegra X-5 adapters directly in the AutoMate 2500 Family Processing Systems
Designed for the clinical lab
- Robust drive system
- Motorized lid-locking system eases loading and unloading
- Surprisingly small footprint saves valuable lab space
- Backed by Beckman Coulter’’s world-class service and support
General Specifications
Depth | 63.4 cm |
Height | 88 cm Door Open |
Width | 49.6 cm |
Power Requirements | 120 VAC, 12A, 60 Hz |
Orientation | Benchtop |
Weight | 77 kg |
Speed Range | 100 to 4700 rpm |
Centrifuge Type | Clinical |
Maximum Capacity | 140 x 13mm blood tubes and 100 x 16mm blood tubes |
Maximum RCF | 4470 xg |
Maximum Speed | 4700 rpm |
Refrigerated | No_ |
Additional Specifications
UNSPSC:
41103903
Height, door closed:
39.0 cm (15.3 in.)
Maximum Heat Dissipation into Room under Steady-State Conditions:
3750 BTU/hr (1.1 kW)
Set Time:
1 min. to 9 hr 59 min, (in 1 min. increments)
Speed Control:
actual rotor speed, ± 50 RPM of set speed
Ambient Temperature Range:
5 to 40°C
Clearances Required:
30 cm (11.8 in.)
Humidity Restrictions:
<80% (noncondensing)
Pollution Degree:
2
Time -- Actual Display:
time remaining in run (timed run ± 1 minute) or ? and elapsed time (continuous run)
Acceleration:
10 acceleration rates
Deceleration:
10 deceleration rates
Installation Category:
II
Speed Display:
actual rotor speed in 100-RPM increments or in RCF (when selected)