This product is Discontinued

Malvern Panalytical - Viscotek HT-GPC

Manufactured by  Malvern Panalytical
Follow this Equipment

Advanced triple detection, using RI, UV, light scattering and a viscometer, for demanding applications.

The HT-GPC system includes a high temperature RI, for use as a concentration detector, a light scattering detector for absolute molecular weight determination and a viscometer for intrinsic viscosity determination and branching. All of these detectors are optionally so the system can be customized to conventional and universal calibration systemsIt also features a temperature controlled column chamber and automated filtering of the sample at high temperature, to ensure safe handling.
Active Questions & AnswersAsk a Question

There are no current Discussions

GPC Gel Permeation Chromatography Service ProvidersView All (17)

Request Support
Documents & Manuals

There are no Documents or Manuals available.

Features of Viscotek HT-GPC
  • Dual piston pump optimized for HT-GPC, with extremely low pulsation and stable eluent flow to improve detector performance.     * Temperature controlled environment for up to 7 columns, injector and up to three detectors. Operates up to 160°C     * Automated sample preparation using the Vortex autoprep/autosampler. Operates up to 170°C with no user intervention, making the system safe, accurate and productive.     * Automated sample filtration reduces the number of preparation steps - only the sample injected on the column is filtered, which avoids bulk filtration and improves lab safety (no manual handling of hot solvents).     * High sensitivity viscometer, with unrivalled baseline stability and the widest application flexibility.     * Digital Inert Transducer (DIT) technology for faster, more sensitive response.     * Unique 7° Low Angle (LALS) detector measures the absolute MW directly without data fitting, extrapolation or correction.     * 90° Right Angle (RALS) detector provides the best signal to noise ratio and the highest sensitivity for small molecules.     * Accurate measurement of the polymer concentration using a refractive index detector.
General Specifications

There are no General Specifications available.

Additional Specifications

Component 1     
Component: Oven/Detector module.     
Temperature Control: 30°C to 160°C all detector cells and columns.     
Column Capacity: Up to 7 standard 30cm GPC/SEC columns.     
Interface: 6 channel digital interface to OmniSEC software.     
Inputs: 2 analog inputs for auxiliary detectors.     
Dimensions: 61cm x 49cm x 56cm (W x D x H)     
     
Component 2     
Component: Vortex Autosampler.     
Capacity: 30 vials, 40mL.     
Temperature Range: 30°C to 170°C.     
Stirring Speed: Variable, 0 - 300 rpm.     
Sample Filtration: Automatic self-cleaning in-line sample filtration.     
Sample Preparation: Optional semi-automated sample preparation.     
Dimensions: 31cm x 31cm x 60cm (W x D x H)     
Weight: 37kg     
     
Component 3     
Component: Degasser.     
Channels/volume: Two channels, 8mL per channel.     
Performance: <0.5ppm oxygen at 0.5mL/min.     
     
Component 4     
Component: Pump.     
Flow rate range: 0.01 to 9.99mL/min.     
Pulsation: Less than 1% (measured with viscometer).     
Pressure Reading: MPa or PSI.     
Soft start/stop: User programmable in mL/min/min.     
     
Detector 1:     
Detector: Light scattering detector.     
Measurement principle: RALS, 90° angle; LALS, 7° angle, with high efficiency optics.     
Light source: Temperature controlled laser diode, 10mW, 660nm     
Cell volume: 18µL     
Data rate: 100Hz collection, DSP with 5Hz to OmniSEC.     
     
Detector 2:     
Detector: Viscometer.     
Measurement principle: 4 capillary differential Wheatstone bridge configuration.     
Transducers: Digital Inert Transducer technology with over-pressure protection.     
Measuring volume: 18µL     
Purging: 3 modes, manual, automatic and programmable.     
Data rate: 100Hz collection, DSP with 5Hz to OmniSEC.     
     
Detector 3:     
Detector: Differential refractive index detector.     
Measurement principle: Deflection     
Light source: LED, 660nm.     
Cell volume: 45° quartz glass, 12µL volume.     
Purging: 3 modes, manual, automatic and programmable.     
Data rate: 100Hz collection, DSP with 5Hz to OmniSEC.     
     
Software:     
OmniSEC, 6 channels at 5Hz, unlimited acquisition time     
     
General     
Laser safety: Class 1.     
Temperature: 22°C - 40°C (Mobile phase (typically Phenol-TCB) must not be allowed to freeze in the system)     
Humidity: 35% - 80% non-condensing.     
Minimum computer specification: 100 GB HD, 2GB RAM, 1.8 GHz processor, Graphics 1024x768, 16-bit color depth     
Operating platforms: Windows XP (SP3), Windows 7 operating system recommended.