Beckman Coulter - LS6000
The Beckman LS 6000 Liquid Scintillation Counter is a high-performing system that offers flexibility of use. This model...
LS 6000 units are equipped with a 32,768 channel multichannel analyzer, delivering a resolution eight times more effective than the standard. Conventional liquid scintillation counters only produce 0.5 keV per channel as opposed to the 0.06 keV per channel from this model. This feature sets the basis for Defects Per Million (DPM) results without quench curves, a precise quench monitor, superior results for multi-label studies and automatic color detection and correction.
Active Questions & AnswersAsk a Question
Won't calibrate
No activity at power up
Scintillation Counters Service ProvidersView All (1)
Documents & Manuals
There are no Documents or Manuals available.
Features of LS6000
- 32,768-Channel Multichannel Analyzer
- Advanced Photomultiplier Design
- Interlocking Lead Shielding
- Windowing User Interface
- User-Selectable Output Format
General Specifications
Depth | 80 cm |
Height | 66 cm |
Width | 91.5 cm |
Power Requirements | 120 / 240V / BTU HR with 50/60 Hz |
Weight | 210 kg |
Additional Specifications
Efficiency:
2H: >/= 60% in a wide window
14C: >/= 95/5 in a wide window
Reproducibility:
+/- 1% Counting Efficiency
Maximum Count Rate
Total Singles:
26 x 10^6 CPM
3H Window:
10 x 10^6 CPM
14C Window:
13 x ^6 CPM
Above 14 C Window :
5 x ^6 CPM
Power Requirements:
120 / 240V / BTU HR with 50/60 Hz
Instrument:
3A / 1.5A / 1230
Temperature:
7.5A / 3.8A / 3060
Inrush Current:
5A at 120V, 2.5A at 240V
Power Failure Recovery:
Systems returns to a disrupted Automatic Count with fully charged battery
Dimensions (W x H x D / Weight)
Instrument:
91.5 x 66 x 80 cm / 210 kg
Monitor:
33 x 34.3 x 35.6 cm / 8.2 kg
Printer:
39 x 10 x 30.5 cm / 5 kg
Ambient Temperature Range:
15 - 35'C
Maximum Relative Humidity:
85%
Ambient Temperature Range:
15'C to 30'C
Maximum Pull Down Time:
8 Hours
Set Point 1:
12'C +3'C
Set Point 2:
15'C +3'C
Set Point 3:
18'C +3'C
Efficiency:
2H: >/= 60% in a wide window
14C: >/= 95/5 in a wide window