Beckman Coulter - Avanti J-26S
The Avanti J-26S series is the only high performance centrifuge system you’ll ever need. Built upon benchmark...
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Updated byMetrology.STL
Centrifuge Service ProvidersView All (22)
Documents & ManualsView All Documents
Features of Avanti J-26S
High Performance Application versatility: can process six liters of fermentor output in less than 10 minutes, deliver a subcellular fraction at 82,000 x g, run 24 microplates at a time, and separate live cell populations with elutriation.
Complete BioSafety* Systems BioCertified** rotors, instruments and labware contain microbiological aerosols, while sterile filters prevent infectious materials from being expelled into the lab environment. Together, they enhance peace of mind for users and lab managers.
Dual System Rotor Recognition Automatically detects which rotor is installed and prevents the rotor from running above its maximum-rated speed—without shutting down the run. All to ensure the safety of your lab.
Friction Reduction System Enables quicker, quieter, more energy-efficient runs with better temperature control, which improves system reliability. High-torque, switched reluctance (SR) drive technology shortens cycle times.
Lightweight J-Lite Rotors Consistently designed to be the most efficient rotors—for reduced run times and high g-force.
Ergonomic Design Lower work surface for easier rotor handling. Foot pedal opens the door and improved door design swings up and out of the way.
Sustainability is Built In Each instrument saves approximately 9.6 pounds of post-consumer materials from the landfill.
General Specifications
Depth | 860 mm |
Height | 860 mm (1490 mm with door open) |
Width | 710 mm |
Orientation | Floor Standing |
Noise Level | 57 dB |
Weight | 290 kg |
Part Number | 41103905 |
Speed Range | 100 to 26000 rpm |
Centrifuge Type | High Speed |
Maximum Capacity | 6 liters |
Maximum RCF | 82000 xg |
Maximum Speed | 26000 rpm |
Additional Specifications
Maximum Heat Dissipation into Room under Steady-State Conditions: 2.0 kW (6,900 BTU/hr)
Refrigeration System: Non-CFC, non-ozone depleting refrigerant
Rotor ID: Automatic
Set Temperature: -10 to +40° C in 1° C increments
Set Time: 1 min to 99 hr 59 min, W2T, or Continuous Hold
Speed Control: ± 10 rpm of set speed or 0.1%, whichever is greater
Single Tube Sample Volume Range: 1.8 - 1,000 mL
Ambient Temperature Range: 16 to 38° C