Tecan - Sunrise
The smart & reliable microplate reader for ELISA assays & kinetics
A set of options such as temperature control up to 42°C, a gradient filter for wavelengths scanning and a barcode scanner allow to customize the Sunrise to your application needs.
The compact space saving design makes the Sunrise ideal for stand-alone use, as well as integration into automated ELISA-systems such as the Freedom EVOlyzer®.
Active Questions & AnswersAsk a Question
lamp low message
Recent Questions & Answers
Updated bydr.wo
Microplate Readers Service ProvidersView All (12)
Documents & ManualsView All Documents
Features of Sunrise
Supports ELISA- & kinetics-, as well as agglutination reading
Selection of user interface: Netbook* or Touchscreen
Selection of eight languages for easy operation
3 year warranty as Tecan´s commitment to quality
Reads plate in only 6 sec. via advanced 12-channel optics
Powered by smart Magellan™ data analysis software
Quick access to your program favourite via start-screen
Advanced curve-fitting & plate-to-plate QC functions
IQ / OQ procedure & documentation
Designed to meet IVDD 98/79/EC & FDA´s 21 CFR part 11**
Netbook not included. ** Using Magellan Tracker software
General Specifications
Shaking | 1 |
Detection Mode | Absorbance |
Shaking Pattern | Linear shaking in 4 different modes |
Wavelength Range | 340 to 750 nm |
Microplate Size | 96-well |
Software | Integrated |
Temperature Control | 1 |
Additional Specifications
Optical system: 12 measurement channels, 1 reference channel
Wavelength selection: Interference filters (slide of 4 to 6 filters), Bandwidth: 10 nm
Measurement range: 340 to 399 nm: 0 to 3.0 OD, 400 to 750 nm: 0 to 4.0 OD
Measurement time:
Single wavelength: 6 seconds
Dual wavelength: 8 seconds
Kinetic repeat time: 5 seconds
Accuracy:
0.0 to 2.0 OD/492 nm: better than ± (1.0% + 0.010 OD)
2.0 to 3.0 OD/492 nm: better than ± (1.5% + 0.010 OD)
Linearity:
340 to 399 nm: 0.0 to 2.0 OD better than ±2%
400 to 750 nm: 0.0 to 2.0 OD better than ±1%
2.0 to 3.0 OD better than ±1.5%
Precision:
0.0 to 2.0 OD/492 nm: better than ± (0.5% + 0.005 OD)
2.0 to 3.0 OD/492 nm: better than ± (1.0% + 0.005 OD)
Resolution: 0.001 OD
Interface: RS 232 C/USB