BINDER - MKFT Series
Environmental simulation chamber for low temperature alternating climate profiles
Active Questions & AnswersAsk a Question
Recent Questions & Answers
Environmental Chambers Service ProvidersView All (20)
Documents & ManualsView All Documents
Features of MKFT Series
- Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
- Temperature range without humidity: -70 °C to 180 °C
- Temperature range with humidity: 10 °C to 95 °C
- Humidity range 10 % to 98 % RH
- MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
- User-friendly LCD screen
- Easy-to-read menu guide
- Integrated electronic chart recorder
- Variety of options for the graphic display of process parameters
- Real-time clock
- Electronically controlled humidification and dehumidification system with capacitive humidity sensor and vapor pressure humidification
- Integrated water storage tank
- Heated viewing window with LED interior lighting
- Programmable condensation protection for test material
- Adjustable ramp function via program editor
- 230 V power socket on the right-side operating panel
- Independent adjustable temperature safety device Class 2
- 4 potential-free relay outputs that can be activated via MCS controller
- Ethernet interface for use with optional GMP/GLP and FDA guideline 21 CFR Part 11 compliant APT-COM™ DataControlSystem software
- Access port Ø 50 mm, left side
- 4 casters (2 with brakes)
- Stainless steel rack included
- BINDER test confirmation
General Specifications
Depth | 1000 mm (incl. cable and door handle) |
Height | 1940 mm (incl. castors) |
Width | 1135 mm |
Weight | 415 kg |
Additional Specifications
Exterior dimensions
Wall clearance rear (mm): 300
Wall clearance side (mm): 260
Viewing window width (mm): 508
Viewing window height (mm): 300
Number of doors (ea.): 1
Interior dimensions
Width (mm): 735
Height (mm): 700
Depth (mm): 443
Interior volume (l): 228
Racks (number standard/max.): 1 / 6
Load per shelf (kg): 30
Permitted total load (kg): 70
Temperature data
Temperature range (°C): -70 - 180
Temperature fluctuation (±K): 0,1 - 0,5
Temperature variation (±K): 0,2 - 1,8
Mean warm-up rate acc. to IEC 60068-3-5 (K/min.): 5,0
Mean cooling rate acc. to IEC 60068-3-5 (K/min.): 4,2
Max. heat compensation up to 25 °C (W): 3000
Climate data
Temperature range (°C): 10 - 95
Temperature fluctuation (± K): 0,1 - 1,5
Humidity range (% RH): 10 - 98
Humidity fluctuation (± % RH): = 2,5
Dew point temperature range (°C): 5 - 94
Max. heat compensation to 25 °C / 90 % RH (W): 400
Electrical data
IP protection class acc. to EN 60529: IP 20
Voltage (± 10%) 50 Hz (V): 400 3N~
Nominal power (kW): 7,5
Energy consumption 25 °C / 60 % RH (Wh/h) 1): 1500
Noise level approx. (dB (A)): 65