ELGA - PURELAB Ultra
Manufactured by ELGA
Ultra Pure Water for Absolute Confidence in Your Results
The PURELAB Ultra is ideal for critical applications such as ultra trace analysis, ICP-MS (Inductively Coupled Plasma Mass Spectrometry) and molecular techniques requiring the ultimate inorganic water purity. This system consistently delivers the highest level of ultra pure water which enables you to concentrate on obtaining accurate results.
Active Questions & AnswersAsk a Question
1Replies
parts needed
Recent Questions & Answers
Updated byJkubicki
Water Purification Service ProvidersView All (6)
Documents & ManualsView All Documents
Features of PURELAB Ultra
- 18.2 MΩ-cm, type I+,I water suitable for critical applications
- PureSure system guarantees your water purity to 18.2 MΩ-cm and beyond
- Delivers up to 2 liters per minute of Type I+,I water
- Full spectrum UV technology to control bacteria levels
- PureSure system ensures that your work is not interrupted
- Real time TOC monitoring reports on TOC every 2 seconds
- Full service and validation support
General Specifications
There are no General Specifications available.
Additional Specifications
Treated Water Specifications
Flowrate: 2.0 l/min max
Inorganics: 18.2 MO-cm
TOC: 1 – 3 ppb¹
Bacteria: <0.1 CFU/ml²
Bacterial endotoxin: <0.001 EU/ml
pH: Effectively neutral
Particles: Ultrafiltration
RNase / DNase: Removed
Labpure cartridge capacity (LC182): Liters at 18.2 MO-cm = 80,000 / µS/cm + ( 2.3 x ppm CO2 )
1 Dependant on feedwater - recommended feed < 50 ppb TOC.
2 With POU filter fitted.
Dimensions and weights
Dimensions: Height 490mm (19.3”), Width 410mm (16.2”), Depth 365mm (14.4”)
Weight: 15.0kg (33.1 lb)
Series Modals
ELGA - PURELAB Ultra Analytic
[+] View Model SpecificationsDepth | 14.4 in |
Flow Rate | 130 L/hr |
Height | 19.3 in |
Width | 16.2 in |
Power Requirements | 100-240 VAC, 50/60 Hz |
Water Purification Method | Other |
Weight | 33.1 lbs |
ASTM Type | ASTM Type I |
TOC | 2 ppb |
Additional Model Specifications | Treated Water Specifications Flowrate: 2.0 I/min max Inorganics: 18.2 MΩ-cm Bacteria: <0.1 CFU/ml² pH: Effectively neutral Particles: 0.05 µm Labpure cartridge capacity (LC182): Liters at 18.2 MΩ-cm = 80,000 / µS/cm + ( 2.3 x ppm CO2 ) 1 Dependant on feedwater - recommended feed < 50 ppb TOC. 2 With POU filter fitted. Feedwater Requirements Source - originally from potable supply, then pre-treated: Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled. Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion. Fouling Index (max): 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds Service Deionization (SDI) - MO-cm: 1 MO-cm minimum resistivity at exhaustion Reverse Osmosis (RO) - µS/cm: Recommended < 30 µS/cm Free chlorine: 0.05 ppm max TOC: Recommended 50 ppb max Carbon dioxide: 30 ppm max Silica: 2 ppm max Particulates: Filtration down to 0.2 micron advisable to protect internal and/or point of use filters Drain requirements (gravity fall with air gap). Maximum during service: Up to 2 l/min Electrical Requirements System voltage: 24V dc Power consumption during recirculation: 60VA Power consumption during dispense: 75VA Fuses: 2 x T6.3 Amp Reservoir level connection: Jack Plug 3.5mm Noise level during recirculation: <40dBA |
ELGA - PURELAB Ultra Bioscience
[+] View Model SpecificationsDepth | 14.4 in |
Flow Rate | 130 L/hr |
Height | 19.3 in |
Width | 16.2 in |
Power Requirements | 100-240 VAC, 50/60 Hz |
Water Purification Method | Other |
Weight | 33.1 lbs |
ASTM Type | ASTM Type I |
TOC | 10 ppb |
Additional Model Specifications | Treated Water Specifications Flowrate: 2.0 L/min max Inorganics: 18.2 MΩ-cm Bacteria: <0.1 CFU/ml² Bacterial endotoxin: <0.001 EU/ml pH: Effectively neutral Particles: Ultrafiltration RNase / DNase: Removed Labpure cartridge capacity (LC182): Liters at 18.2 MΩ-cm = 80,000 / µS/cm + ( 2.3 x ppm CO2 ) 1 Dependant on feedwater - recommended feed < 50 ppb TOC. 2 With POU filter fitted. Feedwater Requirements Source - originally from potable supply, then pre-treated: Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled. Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion. Fouling Index (max): 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds Service Deionization (SDI) - MO-cm: 1 MO-cm minimum resistivity at exhaustion Reverse Osmosis (RO) - µS/cm: Recommended < 30 µS/cm Free chlorine: 0.05 ppm max TOC: Recommended 50 ppb max Carbon dioxide: 30 ppm max Silica: 2 ppm max Particulates: Filtration down to 0.2 micron advisable to protect internal and/or point of use filters Drain requirements (gravity fall with air gap). Maximum during service: Up to 2 l/min Electrical Requirements System voltage: 24V dc Power consumption during recirculation: 60VA Power consumption during dispense: 75VA Fuses: 2 x T6.3 Amp Reservoir level connection: Jack Plug 3.5mm Noise level during recirculation: <40dBA |
ELGA - PURELAB Ultra Genetic
[+] View Model SpecificationsDepth | 14.4 in |
Flow Rate | 130 L/hr |
Height | 19.3 in |
Width | 16.2 in |
Power Requirements | 100-240 VAC, 50/60 Hz |
Water Purification Method | Other |
Weight | 33.1 lbs |
ASTM Type | ASTM Type I |
TOC | 3 ppb |
Additional Model Specifications | Treated Water Specifications Inorganics: 18.2 MΩ-cm Bacteria: <0.1 CFU/ml² Bacterial endotoxin: <0.001 EU/ml pH: Effectively neutral Particles: Ultrafiltration RNase / DNase: Removed Labpure cartridge capacity (LC182): Liters at 18.2 MΩ-cm = 80,000 / µS/cm + ( 2.3 x ppm CO2 ) 1 Dependant on feedwater - recommended feed < 50 ppb TOC. 2 With POU filter fitted. Feedwater Requirements Source - originally from potable supply, then pre-treated: Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled. Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion. Fouling Index (max): 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds Service Deionization (SDI) - MO-cm: 1 MO-cm minimum resistivity at exhaustion Reverse Osmosis (RO) - µS/cm: Recommended < 30 µS/cm Free chlorine: 0.05 ppm max TOC: Recommended 50 ppb max Carbon dioxide: 30 ppm max Silica: 2 ppm max Particulates: Filtration down to 0.2 micron advisable to protect internal and/or point of use filters Drain requirements (gravity fall with air gap). Maximum during service: Up to 2 l/min Electrical Requirements System voltage: 24V dc Power consumption during recirculation: 60VA Power consumption during dispense: 75VA Fuses: 2 x T6.3 Amp Reservoir level connection: Jack Plug 3.5mm Noise level during recirculation: <40dBA |
ELGA - PURELAB Ultra Ionic
[+] View Model SpecificationsDepth | 14.4 in |
Flow Rate | 130 L/hr |
Height | 19.3 in |
Width | 16.2 in |
Power Requirements | 100-240 VAC, 50/60 Ha |
Water Purification Method | Other |
Weight | 33.1 lbs |
ASTM Type | ASTM Type I |
TOC | 10 ppb |
Additional Model Specifications | Treated Water Specifications Flowrate: 2.0 L/min max Inorganics: 18.2 MΩ-cm Bacteria: <0.1 CFU/ml² pH: Effectively neutral Particles: 0.05 µm Labpure cartridge capacity (LC182): Liters at 18.2 MΩ-cm = 80,000 / µS/cm + ( 2.3 x ppm CO2 ) 1 Dependant on feedwater - recommended feed < 50 ppb TOC. 2 With POU filter fitted. Feedwater Requirements Source - originally from potable supply, then pre-treated: Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled. Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion. Fouling Index (max): 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds Service Deionization (SDI) - MO-cm: 1 MO-cm minimum resistivity at exhaustion Reverse Osmosis (RO) - µS/cm: Recommended < 30 µS/cm Free chlorine: 0.05 ppm max TOC: Recommended 50 ppb max Carbon dioxide: 30 ppm max Silica: 2 ppm max Particulates: Filtration down to 0.2 micron advisable to protect internal and/or point of use filters Drain requirements (gravity fall with air gap). Maximum during service: Up to 2 l/min Electrical Requirements System voltage: 24V dc Power consumption during recirculation: 60VA Power consumption during dispense: 75VA Fuses: 2 x T6.3 Amp Reservoir level connection: Jack Plug 3.5mm Noise level during recirculation: <40dBA |
ELGA - PURELAB Ultra Scientific
[+] View Model SpecificationsDepth | 14.4 in |
Flow Rate | 130 L/hr |
Height | 19.3 in |
Width | 16.2 in |
Power Requirements | 100-240 VAC, 50/60 Hz |
Water Purification Method | Other |
Weight | 33.1 lbs |
ASTM Type | ASTM Type I |
TOC | 2 ppb |
Additional Model Specifications | Treated Water Specifications Flowrate: 2.0 L/min max Inorganics: 18.2 MΩ-cm Bacteria: <1 CFU/ml² pH: Effectively neutral Particles: 0.02 µm² Labpure cartridge capacity (LC182): Liters at 18.2 MΩ-cm = 80,000 / µS/cm + ( 2.3 x ppm CO2 ) 1 Dependant on feedwater - recommended feed < 50 ppb TOC. 2 With POU filter fitted. Feedwater Requirements Source - originally from potable supply, then pre-treated: Preferably reverse osmosis (RO) or filtered service deionisation (SDI) or distilled. Note: mixed bed or twin bed deionised supplies should be cation limited at exhaustion. Fouling Index (max): 1 for all models. A 0.2 micron membrane prefilter is recommended for all non-RO feeds Service Deionization (SDI) - MO-cm: 1 MO-cm minimum resistivity at exhaustion Reverse Osmosis (RO) - µS/cm: Recommended < 30 µS/cm Free chlorine: 0.05 ppm max TOC: Recommended 50 ppb max Carbon dioxide: 30 ppm max Silica: 2 ppm max Particulates: Filtration down to 0.2 micron advisable to protect internal and/or point of use filters Drain requirements (gravity fall with air gap). Maximum during service: Up to 2 l/min Electrical Requirements System voltage: 24V dc Power consumption during recirculation: 60VA Power consumption during dispense: 75VA Fuses: 2 x T6.3 Amp Reservoir level connection: Jack Plug 3.5mm Noise level during recirculation: <40dBA |