Milton Roy - Spectronic 501/601
Manufactured by Milton Roy
Spectronic 501/601 Spectrophotometer
The Milton Roy 501/601 Spectrophotometer is a simple instrument designed for quick and easy absorbency or transmittance measurements at single wavelengths. The 501 has a range of 325 to 999nm, while the 601 includes the UV range, (195-999nm), with the addition of a Deuterium lamp. These units come with a variety of options such as a heated or cooled sample holder, a recorder output, and an RS232 output. The units are designed for single wavelength measurements but a scanning program could be written via the RS232 communications port. All necessary information is given in the operator’s manual in section 9. Though the instrument could not be very accurate between wavelengths far apart, it could be effective for short scans.
Operation is very basic. Enter the desired wavelength, zero the blank and read the sample. Results can be printed out to a printer but it needs to accept serial data and at the set parameters of the unit.
Operation is very basic. Enter the desired wavelength, zero the blank and read the sample. Results can be printed out to a printer but it needs to accept serial data and at the set parameters of the unit.
Active Questions & AnswersAsk a Question
6Replies
Calibration test problem
3Replies
Technician
UV/VIS Spectrophotometers Service ProvidersView All (32)
Documents & ManualsView All Documents
Features
There are no Features available.
General Specifications
There are no General Specifications available.
Additional Specifications
Series Modals
Milton Roy - 501
[+] View Model SpecificationsDepth | 14.3 in |
Height | 7.75 in |
Width | 17.5 in |
Power Requirements | 100/115 and 220/240 V. ± 10%; 50/60 Hz, ± 2 Hz; 2.0 A |
Photometric Range | Absorbency: -0.1 to 3.0 A, Transmittance: 0.0 to 125% T, concentration: -9999 to + 9999 |
Weight | 35 lbs |
Wavelength Range | 325 to 999 nm |
Light Source | Tungsten-Halogen |
Optics | Single Beam |
Photodetector | Photodiode |
Photometric Noise | Less than 0.004 A near 2 A nm, Less than 0.0015 A near 0 A |
Sample Compartment | Dimensions: 10.2 cm (4 in) High x 10.6 cm (4.2 in) Wide x 17.1 cm (6.5 in) Deep |
Stray Light | <= 0.05% T at 220nm and 340 nm |
Additional Model Specifications | Wavelength accuracy: ± 1.0 nm Wavelength precision: Better than 0.5 nm Spectral slit width: <= 5 nm Photometric accuracy: ± 0.005 A near 1.0 A Photometric stability: ± 0.003 A per hour near 0 A Storage Conditions: -40 to 105 F; -40 to 40.5 C; relative humidity not to exceed 60% (Allow unit to adjust to room temperature for 24 hours when taking it out of storage.) |
Milton Roy - 601
[+] View Model SpecificationsDepth | 14.3 in |
Height | 7.75 in |
Width | 17.5 in |
Power Requirements | 100/115 and 220/240 V. ± 10%; 50/60 Hz, ± 2 Hz; 2.0 A |
Photometric Range | Absorbency: -0.1 to 3.0 A, Transmittance: 0.0 to 125% T, concentration: -9999 to + 9999 |
Weight | 35 lbs |
Wavelength Range | 195 to 999 nm |
Light Source | Deuterium |
Optics | Single Beam |
Photodetector | Photodiode |
Photometric Noise | Less than 0.004 A near 2 A nm, Less than 0.0015 A near 0 A |
Sample Compartment | Dimensions: 10.2 cm (4 in) High x 10.6 cm (4.2 in) Wide x 17.1 cm (6.5 in) Deep |
Stray Light | <= 0.05% T at 220nm and 340 nm |
Additional Model Specifications | Wavelength accuracy: ± 1.0 nm Wavelength precision: Better than 0.5 nm Spectral slit width: <= 5 nm Photometric accuracy: ± 0.005 A near 1.0 A Photometric stability: ± 0.003 A per hour near 0 A Storage Conditions: -40 to 105 F; -40 to 40.5 C; relative humidity not to exceed 60% (Allow unit to adjust to room temperature for 24 hours when taking it out of storage.) |